Reddy V R Sudha, M E Sumathi, Gowda Y C Beere, Suhail S Mohamed
Associate Professor, Department of Paediatrics, Sri Devaraj Urs Medical College , Tamaka, Kolar, Karnataka, India .
Associate Professor, Department of Biochemistry, Sri Devaraj Urs Medical College , Tamaka, Kolar, Karnataka, India .
J Clin Diagn Res. 2014 Jul;8(7):PC10-3. doi: 10.7860/JCDR/2014/8666.4538. Epub 2014 Jul 20.
Neonatal hypoglycemia is the most common carbohydrate metabolic disturbance seen in case of neonates and especially in preterm neonates. Accurate and rapid determination of hypoglycemia and its prompt treatment is of utmost importance to decrease morbidity and mortality of neonates.
To estimate blood glucose in neonates and test the efficacy of HemoCue Glucose 201+ analyser and B Braun Glucometer by comparing with centralised laboratory testing. To compare the blood glucose in capillary and venous blood samples of neonates.
Hospital setting; Comparative Study Materials and Methods: After obtaining informed consent, all neonates admitted to Neonatal Intensive Care Unit (NICU) were screened for blood sugar. Capillary and venous blood glucose was estimated employing HemoCue Glucose 201+ analyser and B Braun Glucometer. Simultaneously, the same venous sample was collected in fluoride tube and sent to central clinical biochemistry laboratory for glucose estimation. When anaemia or polycythemia was clinically suspected the same venous sample was sent for estimation of Hematocrit (Hct).
Comparison of blood glucose concentration of B Braun glucometer, HemoCue Glucose 201+ analyser and centralis/ed plasma glucose levels was done by using students test. All the statistical analysis were done using software SPSS 6 version.
Mean values of blood glucose (100.2 + 48.4) with B Braun glucometer was significantly higher (p=0.003) when compared to plasma glucose values (76.95 + 45.99) estimated in central laboratory and HemoCue glucose 201+ analyser (82.9 + 51.4). HemoCue glucose 201+ analyser did not show significant difference (p=0.463) with central laboratory testing. There was no significant difference between the capillary and venous sample estimated in both the instruments. Estimation with HemoCue glucose 201+ analyser correlated well with central laboratory testing in neonates with blood glucose <55mg/dl,
We conclude that HemoCue glucose 201+ analyser appears to be a suitable point of care (POC) blood glucose measurement device in neonates on both capillary and venous blood samples, as it showed a good correlation with central laboratory values without significant interference from Hct.
新生儿低血糖是新生儿尤其是早产儿中最常见的碳水化合物代谢紊乱。准确、快速地测定低血糖并及时治疗对于降低新生儿的发病率和死亡率至关重要。
评估新生儿血糖水平,并通过与中心实验室检测结果对比,检验HemoCue Glucose 201+分析仪和贝朗血糖仪的效能。比较新生儿毛细血管血和静脉血样本中的血糖水平。
医院环境;对比研究。材料与方法:在获得知情同意后,对入住新生儿重症监护病房(NICU)的所有新生儿进行血糖筛查。使用HemoCue Glucose 201+分析仪和贝朗血糖仪测定毛细血管血和静脉血中的血糖。同时,将同一静脉血样本采集到含氟试管中,送至中心临床生化实验室进行血糖测定。当临床怀疑有贫血或红细胞增多症时,将同一静脉血样本送去测定血细胞比容(Hct)。
采用学生检验对贝朗血糖仪、HemoCue Glucose 201+分析仪和中心血浆葡萄糖水平的血糖浓度进行比较。所有统计分析均使用SPSS 6版本软件完成。
贝朗血糖仪测得的血糖平均值(100.2 + 48.4)与中心实验室测得的血浆葡萄糖值(76.95 + 45.99)以及HemoCue葡萄糖201+分析仪测得的值(82.9 + 51.4)相比,显著更高(p = 0.003)。HemoCue葡萄糖201+分析仪与中心实验室检测结果无显著差异(p = 0.463)。两种仪器测定的毛细血管血和静脉血样本之间无显著差异。对于血糖<55mg/dl的新生儿,HemoCue葡萄糖201+分析仪的测定结果与中心实验室检测结果相关性良好。
我们得出结论,HemoCue葡萄糖201+分析仪似乎是一种适用于新生儿毛细血管血和静脉血样本的即时检测(POC)血糖测量设备,因为它与中心实验室值显示出良好的相关性,且不受Hct的显著干扰。