Matlach Juliane, Wagner Martin, Malzahn Uwe, Göbel Winfried
Department of Ophthalmology, University of Würzburg, Germany.
Institute of Clinical Epidemiology and Biometry, University of Würzburg, Germany Comprehensive Heart Failure Center, University of Würzburg, Germany.
Invest Ophthalmol Vis Sci. 2014 Sep 16;55(10):6536-46. doi: 10.1167/iovs.14-15072.
To compare measurement of macular inner retina and peripapillary retinal nerve fiber layer (pRNFL) thickness using two spectral-domain optical coherence tomography (SD-OCT) devices in glaucoma patients, patients with ocular hypertension, idiopathic and atypical Parkinson disease, and healthy controls.
A total of 171 eyes of 146 participants underwent successful pRNFL and macular scanning and automated measurement of ganglion cell layer+inner plexiform layer (GCL-IPL) using Cirrus HD-OCT or retinal nerve fiber layer+GCL-IPL (RNFL-GCL-IPL) using RTVue-100. Macular RNFL was added to the GCL-IPL thickness measured by Cirrus and was compared to the RNFL-GCL-IPL thickness of the RTVue in the corresponding Cirrus sectors. Intraclass correlation coefficient (ICC) was calculated to determine repeatability of three consecutive measurements; ICC and Bland-Altman analysis to assess agreement between OCTs; Pearson's correlation coefficient to assess strength of linear correlation.
Repeatability of average macular RNFL-GCL-IPL thickness measurement was excellent with an ICC of 0.998 for Cirrus and 0.995 for RTVue. Repeatability was also good for average pRNFL thickness measurements. Both instruments demonstrated a good consistency in measurements with ICC values ranging from 0.844 to 0.862 for macular RNFL-GCL-IPL and 0.718 to 0.958 for pRNFL thickness.
Measurement of pRNFL and macular RNFL-GCL-IPL thickness has a high degree of repeatability for both OCT devices. Despite a high correlation between measurements of the two OCT devices and fair to excellent ICC values representing a high consistency in the measurements of the two devices, RTVue measured a thicker macular RNFL-GCL-IPL and pRNFL thickness.
使用两种光谱域光学相干断层扫描(SD-OCT)设备,比较青光眼患者、高眼压症患者、特发性和非典型帕金森病患者以及健康对照者黄斑内层视网膜和视乳头周围视网膜神经纤维层(pRNFL)厚度的测量结果。
146名参与者的171只眼睛成功进行了pRNFL和黄斑扫描,并使用Cirrus HD-OCT自动测量神经节细胞层+内网状层(GCL-IPL),或使用RTVue-100自动测量视网膜神经纤维层+GCL-IPL(RNFL-GCL-IPL)。将Cirrus测量的GCL-IPL厚度加上黄斑RNFL,并与相应Cirrus扇区中RTVue的RNFL-GCL-IPL厚度进行比较。计算组内相关系数(ICC)以确定连续三次测量的可重复性;使用ICC和Bland-Altman分析评估两种OCT之间的一致性;使用Pearson相关系数评估线性相关强度。
Cirrus测量黄斑RNFL-GCL-IPL平均厚度的可重复性极佳,ICC为0.998,RTVue为0.995。pRNFL平均厚度测量的可重复性也很好。两种仪器在测量方面均显示出良好的一致性,黄斑RNFL-GCL-IPL的ICC值范围为0.844至0.862,pRNFL厚度的ICC值范围为0.718至0.958。
两种OCT设备测量pRNFL和黄斑RNFL-GCL-IPL厚度均具有高度可重复性。尽管两种OCT设备的测量结果之间具有高度相关性,且ICC值良好至极佳,表明两种设备的测量具有高度一致性,但RTVue测量的黄斑RNFL-GCL-IPL和pRNFL厚度更厚。