A new approach to the noninvasive measurement of cardiac output using an annular array Doppler technique--II. Practical implementation and results.

作者信息

Evans J M, Skidmore R, Baker J D, Wells P N

机构信息

Department of Medical Physics, Bristol General Hospital, England.

出版信息

Ultrasound Med Biol. 1989;15(3):179-87. doi: 10.1016/0301-5629(89)90062-8.

Abstract

An experimental Doppler flowmeter system has been developed which can noninvasively measure blood flow volume rate in a vessel. It is based on the attenuation compensated technique and does not require knowledge of the vessel size or beam-vessel angle. In vitro results have shown that the measurement of volume flow rate is independent of vessel angle to within +/- 4%, and independent of vessel diameter to within +/- 5%. Flow rate linearity is better than +/- 3%. A good comparison has been obtained, in vivo, of aortic diameters measured by an imaging system and with this flowmeter; the r value was 0.98. The noninvasive measurement of cardiac output using this flowmeter has been compared with conventional dilution techniques in 54 patients, with a resulting correlation coefficient of r = 0.96.

摘要

已开发出一种实验性多普勒流量计系统,它能够无创测量血管中的血流量。该系统基于衰减补偿技术,无需了解血管大小或波束与血管的夹角。体外实验结果表明,体积流量的测量与血管夹角的偏差在±4%以内,与血管直径的偏差在±5%以内。流量线性度优于±3%。在体内,通过成像系统测量的主动脉直径与使用该流量计测量的结果进行了良好的对比;相关系数r为0.98。在54例患者中,使用该流量计对心输出量进行的无创测量与传统稀释技术进行了比较,结果相关系数r = 0.96。