Mater Sci Eng C Mater Biol Appl. 2014 Nov;44:240-5. doi: 10.1016/j.msec.2014.08.004. Epub 2014 Aug 7.
A new on-line separation/preconcentration system was developed for the determination of Cu(II) ions by flame atomic absorption spectrometry in urine samples. A newly synthesized chelating resin, by anchoring eriochrome blue black R reagent to Amberlite XAD-16 resin, was used as a packing material for the selective separation/preconcentration of Cu(II) ions. The influence of the parameters on the determination of Cu(II) ions such as pH of sample solution, amount of the resin, eluent type, interfering ions and flow variables was studied. The detection limit of the method was 1.0 μg L(-1) while precision was 2.3% (n=15) at 50 μg L(-1) Cu(II) level. The adsorption capacity of the resin was 217 μg g(-1) Cu(II). The accuracy of the method was proven using TMDA-64 standard lake water and synthetic urine sample. The developed method has been applied successfully to the determination of copper in urine with satisfactory results.
开发了一种新的在线分离/预浓缩系统,用于通过火焰原子吸收光谱法测定尿液样本中的铜(II)离子。一种新合成的螯合树脂,通过将铬蓝黑R试剂锚定到Amberlite XAD - 16树脂上,用作选择性分离/预浓缩铜(II)离子的填充材料。研究了诸如样品溶液的pH值、树脂用量、洗脱液类型、干扰离子和流量变量等参数对铜(II)离子测定的影响。该方法的检测限为1.0 μg L(-1),在50 μg L(-1)铜(II)水平下精密度为2.3%(n = 15)。树脂对铜(II)的吸附容量为217 μg g(-1)。使用TMDA - 64标准湖水和合成尿液样本验证了该方法的准确性。所开发的方法已成功应用于尿液中铜的测定,结果令人满意。