Dai Cuixia, Fan Shanhui, Chai Xinyu, Li Yan, Ren Qiushi, Xi Peng, Zhou Chuanqing
Appl Opt. 2014 Aug 20;53(24):5375-9. doi: 10.1364/AO.53.005375.
We have demonstrated a dual-channel multiplexing spectral-domain optical-coherence tomography (SD-OCT) system based on a 3×3 fiber coupler for extended imaging range of whole human eye depth, with a single light source and spectrometer. OCT images of anterior segments of a human eye were sequentially performed and constructed to demonstrate an extended depth range as large as 15 mm in air. A good quality OCT image of the whole anterior segment of an eye was present. Furthermore, whole eye segmental imaging was performed and ocular distances were calculated to show the validation of the system for whole eye morphological measurement.
我们展示了一种基于3×3光纤耦合器的双通道复用光谱域光学相干断层扫描(SD-OCT)系统,该系统采用单一光源和光谱仪,可扩展整个人眼深度的成像范围。对人眼前节进行了OCT图像的顺序采集和构建,以展示在空气中高达15毫米的扩展深度范围。获得了质量良好的整个人眼前节OCT图像。此外,还进行了全眼分段成像并计算了眼内距离,以验证该系统用于全眼形态测量的有效性。