Klemetsen O, Birkelund Y, Maccarini P F, Stauffer P, Jacobsen S K
University of Tromso, Norway.
Duke University Medical Center, USA.
Prog Electromagn Res Symp. 2010;2010:932-936.
We have investigated the possibility of building a Dicke radiometer that is inexpensive, small-sized, stable, high sensitivity and consists of readily available microwave components. The selected frequency band is at 3-4 GHz and can be used for breast cancer detection, with sufficient spatial resolution. We have found microwave components that are small (< 5mm × 5 mm) and provide sufficient sensitivity. We have built two different Dicke radiometers: One is of conventional design with Dicke switch at front end to select antenna or noise rererence and the other with a low noise amplifier before the Dicke Switch. We have tested this concept with simulations and built prototypes. The two designs provide a gain of approximately 50 dB, and bandwidth of about 500 MHz. One of the designs has a stability > 1 and the other design provide instability < 1 for a part of the pass band. The prototypes are tested for sensitivity after calibration in two different known temperature waterbaths. The results show that the design with the low noise amplifier before the Dicke switch has 36% higher sensitivity than the other design with Dicke switch in front.
我们研究了制造一种价格低廉、体积小巧、性能稳定、灵敏度高且由易于获取的微波组件组成的狄克辐射计的可能性。所选频段为3 - 4 GHz,可用于乳腺癌检测,具有足够的空间分辨率。我们找到了尺寸小(<5mm×5mm)且灵敏度足够的微波组件。我们制造了两种不同的狄克辐射计:一种是传统设计,前端有狄克开关以选择天线或噪声参考;另一种在狄克开关之前有一个低噪声放大器。我们通过仿真测试了这个概念并制造了原型。这两种设计的增益约为50 dB,带宽约为500 MHz。其中一种设计的稳定性>1,另一种设计在部分通带内的不稳定性<1。在两个不同的已知温度水浴中校准后,对原型进行了灵敏度测试。结果表明,在狄克开关之前有低噪声放大器的设计比前端有狄克开关的另一种设计灵敏度高36%。