Validity of an activity monitor in young people with cerebral palsy gross motor function classification system level I.

作者信息

O' Donoghue Deirdre, Kennedy Norelee

机构信息

Central Remedial Clinic, Clondalkin, Dublin, Ireland.

出版信息

Physiol Meas. 2014 Nov;35(11):2307-18. doi: 10.1088/0967-3334/35/11/2307. Epub 2014 Oct 23.

Abstract

The activPAL™ activity monitor has potential for use in youth with Cerebral Palsy (CP) as it has demonstrated acceptable validity for the assessment of sedentary and physical activity in other populations. This study determined the validity of the activPAL™ activity monitor for the measurement of sitting, standing, walking time, transitions and step count for both legs in young people with hemiplegic and asymmetric diplegic CP. Seventeen participants with CP Gross Motor Function Classification System level I completed two video recorded test protocols that involved wearing an activPAL™ activity monitor on alternate legs. Agreement between observed video recorded data and activPAL™ activity monitor data was assessed using the Bland and Altman (BA) method and intraclass correlation coefficients (ICC 3,1). There was perfect agreement for transitions and high agreement for sitting (BA mean differences (MD): -1.8 and -1.8 s; ICCs: 0.49 and 0.95) standing (MD: 0.8 and 0.1 s; ICCs: 0.59 and 0.98) walking (MD: 1 and 1.1 s; ICCs: 0.99 and 0.94) timings and low agreement for step count (MD: 4.1 and 2.8 steps; ICCs: 0.96 and 0.95) for both legs. This study found clinically acceptable agreement with direct observation for all activPAL™ activity monitor functions, except for step count measurement with respect to the range of measurement values obtained for both legs in this study population.

摘要

activPAL™ 活动监测仪有潜力应用于患有脑瘫(CP)的青少年,因为它已证明在评估其他人群的久坐和身体活动方面具有可接受的效度。本研究确定了activPAL™ 活动监测仪在测量偏瘫和不对称双瘫型脑瘫青少年双腿的坐姿、站姿、行走时间、姿势转换和步数方面的效度。17名脑瘫粗大运动功能分类系统为I级的参与者完成了两个视频记录测试方案,其中包括在双腿上交替佩戴activPAL™ 活动监测仪。使用Bland和Altman(BA)方法以及组内相关系数(ICC 3,1)评估观察到的视频记录数据与activPAL™ 活动监测仪数据之间的一致性。姿势转换方面一致性极佳,坐姿方面一致性较高(BA平均差异(MD):-1.8和-1.8秒;ICC:0.49和0.95),站姿方面(MD:0.8和0.1秒;ICC:0.59和0.98)、行走时间方面(MD:1和1.1秒;ICC:0.99和0.94)一致性较高,双腿步数方面一致性较低(MD:4.1和2.8步;ICC:0.96和0.95)。本研究发现,对于activPAL™ 活动监测仪的所有功能,除了本研究人群中双腿获得的测量值范围内的步数测量外,与直接观察法在临床上具有可接受的一致性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验