Krasny Lukas, Benada Oldřich, Strnadova Marcela, Lemr Karel, Havlicek Vladimir
Institute of Microbiology, v.v.i., Videnska 1083, 14220, Prague 4, Czech Republic.
Anal Bioanal Chem. 2015 Mar;407(8):2141-7. doi: 10.1007/s00216-014-8294-6. Epub 2014 Nov 1.
Nanostructure-assisted laser desorption/ionization (NALDI) has been recognized as a powerful matrix-free mass spectrometry tool ideal for imaging of small molecules. In this report, the NALDI approach was compared with matrix-assisted laser desorption/ionization (MALDI) mass spectrometry in terms of sensitivity, reproducibility, and lateral resolution, which can be achieved in mass spectrometry imaging (MSI) experiments using a Nd:YAG laser. Scanning electron microscopy was used for surface topology analysis and evaluation of a putative surface-enhanced sensitivity effect, which was observed upon reduction of the laser focus. NALDI was identified as a more reproducible technique lacking MSI artifacts arising from distant tissue removal known from MALDI oversampling.
纳米结构辅助激光解吸/电离(NALDI)已被公认为是一种强大的无基质质谱工具,非常适合小分子成像。在本报告中,将NALDI方法与基质辅助激光解吸/电离(MALDI)质谱在灵敏度、重现性和横向分辨率方面进行了比较,这些指标可在使用Nd:YAG激光的质谱成像(MSI)实验中实现。扫描电子显微镜用于表面拓扑分析和对假定的表面增强灵敏度效应的评估,该效应在激光聚焦减小时被观察到。NALDI被确定为一种更具重现性的技术,不存在MALDI过采样中因远距离组织去除而产生的MSI伪像。