Suppr超能文献

改进的热释光剂量测定法对¹²⁵I和¹⁰³Pd近距离治疗种子剂量率常数测定的影响

Effect of improved TLD dosimetry on the determination of dose rate constants for (125)I and (103)Pd brachytherapy seeds.

作者信息

Rodriguez M, Rogers D W O

机构信息

Carleton Laboratory for Radiotherapy Physics, Carleton University, Ottawa, Ontario K1S 5B6, Canada and Princess Margaret Hospital, Toronto, Ontario M5G 2M9, Canada.

Carleton Laboratory for Radiotherapy Physics, Carleton University, Ottawa, Ontario K1S 5B6, Canada.

出版信息

Med Phys. 2014 Nov;41(11):114301. doi: 10.1118/1.4895003.

Abstract

PURPOSE

To more accurately account for the relative intrinsic energy dependence and relative absorbed-dose energy dependence of TLDs when used to measure dose rate constants (DRCs) for (125)I and (103)Pd brachytherapy seeds, to thereby establish revised "measured values" for all seeds and compare the revised values with Monte Carlo and consensus values.

METHODS

The relative absorbed-dose energy dependence, f(rel), for TLDs and the phantom correction, Pphant, are calculated for (125)I and (103)Pd seeds using the EGSnrc BrachyDose and DOSXYZnrc codes. The original energy dependence and phantom corrections applied to DRC measurements are replaced by calculated (f(rel))(-1) and Pphant values for 24 different seed models. By comparing the modified measured DRCs to the MC values, an appropriate relative intrinsic energy dependence, kbq (rel), is determined. The new Pphant values and relative absorbed-dose sensitivities, SAD (rel), calculated as the product of (f(rel))(-1) and (kbq (rel))(-1), are used to individually revise the measured DRCs for comparison with Monte Carlo calculated values and TG-43U1 or TG-43U1S1 consensus values.

RESULTS

In general, f(rel) is sensitive to the energy spectra and models of the brachytherapy seeds. Values may vary up to 8.4% among (125)I and (103)Pd seed models and common TLD shapes. Pphant values depend primarily on the isotope used. Deduced (kbq (rel))(-1) values are 1.074 ± 0.015 and 1.084 ± 0.026 for (125)I and (103)Pd seeds, respectively. For (1 mm)(3) chips, this implies an overall absorbed-dose sensitivity relative to (60)Co or 6 MV calibrations of 1.51 ± 1% and 1.47 ± 2% for (125)I and (103)Pd seeds, respectively, as opposed to the widely used value of 1.41. Values of Pphant calculated here have much lower statistical uncertainties than literature values, but systematic uncertainties from density and composition uncertainties are significant. Using these revised values with the literature's DRC measurements, the average discrepancies between revised measured values and Monte Carlo values are 1.2% and 0.2% for (125)I and (103)Pd seeds, respectively, compared to average discrepancies for the original measured values of 4.8%. On average, the revised measured values are 4.3% and 5.9% lower than the original measured values for (103)Pd and (125)I seeds, respectively. The average of revised DRCs and Monte Carlo values is 3.8% and 2.8% lower for (125)I and (103)Pd seeds, respectively, than the consensus values in TG-43U1 or TG-43U1S1.

CONCLUSIONS

This work shows that f(rel) is TLD shape and seed model dependent suggesting a need to update the generalized energy response dependence, i.e., relative absorbed-dose sensitivity, measured 25 years ago and applied often to DRC measurements of (125)I and (103)Pd brachytherapy seeds. The intrinsic energy dependence for LiF TLDs deduced here is consistent with previous dosimetry studies and emphasizes the need to revise the DRC consensus values reported by TG-43U1 or TG-43U1S1.

摘要

目的

在使用热释光剂量计(TLD)测量¹²⁵I和¹⁰³Pd近距离治疗种子源的剂量率常数(DRC)时,更准确地考虑TLD的相对固有能量依赖性和相对吸收剂量能量依赖性,从而为所有种子源确定修订后的“测量值”,并将修订后的值与蒙特卡罗计算值和共识值进行比较。

方法

使用EGSnrc BrachyDose和DOSXYZnrc代码,针对¹²⁵I和¹⁰³Pd种子源计算TLD的相对吸收剂量能量依赖性f(rel)和体模校正Pphant。应用于DRC测量的原始能量依赖性和体模校正,被24种不同种子源模型的计算值(f(rel))⁻¹和Pphant值所取代。通过将修正后的测量DRC与蒙特卡罗值进行比较,确定适当的相对固有能量依赖性kbq(rel)。新的Pphant值和相对吸收剂量灵敏度SAD(rel)(计算为(f(rel))⁻¹与(kbq(rel))⁻¹的乘积),用于分别修正测量的DRC,以便与蒙特卡罗计算值以及TG - 43U1或TG - 43U1S1共识值进行比较。

结果

一般来说,f(rel)对近距离治疗种子源的能谱和模型敏感。在¹²⁵I和¹⁰³Pd种子源模型以及常见的TLD形状之间,其值可能相差高达8.4%。Pphant值主要取决于所使用的同位素。对于¹²⁵I和¹⁰³Pd种子源,推导得出的(kbq(rel))⁻¹值分别为1.074 ± 0.015和1.084 ± 0.026。对于(1 mm)³的芯片,这意味着相对于⁶⁰Co或6 MV校准,¹²⁵I和¹⁰³Pd种子源的总体吸收剂量灵敏度分别为1.51 ± 1%和1.47 ± 2%,而不是广泛使用的1.41。此处计算的Pphant值的统计不确定性远低于文献值,但密度和成分不确定性带来的系统不确定性很大。将这些修订值与文献中的DRC测量值一起使用时,¹²⁵I和¹⁰³Pd种子源修订后的测量值与蒙特卡罗值之间的平均差异分别为1.2%和0.2%,而原始测量值的平均差异为4.8%。平均而言,¹⁰³Pd和¹²⁵I种子源修订后的测量值分别比原始测量值低4.3%和5.9%。¹²⁵I和¹⁰³Pd种子源修订后的DRC与蒙特卡罗值的平均值分别比TG - 43U1或TG - 43U1S1中的共识值低3.8%和2.8%。

结论

这项工作表明,f(rel)取决于TLD形状和种子源模型,这表明需要更新25年前测量并经常应用于¹²⁵I和¹⁰³Pd近距离治疗种子源DRC测量的广义能量响应依赖性,即相对吸收剂量灵敏度。此处推导的LiF TLD的固有能量依赖性与先前的剂量学研究一致,并强调需要修订TG - 43U1或TG - 43U1S1报告的DRC共识值。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验