UltraFast Doppler ultrasonography for hepatic vessels of liver recipients: preliminary experiences.

机构信息

Department of Radiology, SMG-SNU Boramae Medical Center, Seoul National University Hospital, Seoul, Korea.

Department of Radiology, Seoul National University Hospital, Seoul, Korea ; Department of Institute of Radiation Medicine, Seoul National University Hospital, Seoul, Korea.

出版信息

Ultrasonography. 2015 Jan;34(1):58-65. doi: 10.14366/usg.14038. Epub 2014 Oct 4.

Abstract

PURPOSE

The purpose of this study was to investigate the value of UltraFast Doppler ultrasonography (US) for evaluating hepatic vessels in liver recipients.

METHODS

Thirty-nine liver Doppler US sessions were conducted in 20 liver recipients. Each session consisted of UltraFast and conventional liver Doppler US in a random order. We compared the velocities and phasicities of the hepatic vessels, duration of each Doppler study, occurrence of technical failures, and differences in clinical decisions.

RESULTS

The velocities and resistive index values of hepatic vessels showed a strong positive correlation between the two Doppler studies (mean R=0.806; range, 0.710 to 0.924). The phasicities of the hepatic vessels were the same in both Doppler US exams. With respect to the duration of the Doppler US exam, there was no significant difference between the UltraFast (251±99 seconds) and conventional (231±117 seconds) Doppler studies (P=0.306). In five poor breath-holders, in whom the duration of conventional Doppler US was longer, UltraFast Doppler US (272±157 seconds) required a shorter time than conventional Doppler US (381±133 seconds; P=0.005). There was no difference between the two techniques with respect to technical failures and clinical decisions.

CONCLUSION

UltraFast Doppler US is clinically equivalent to conventional Doppler US with advantages for poor breath-holders during the post-liver transplantation work-up.

摘要

目的

本研究旨在探讨 UltraFast 多普勒超声(US)在评估肝移植受者肝血管中的价值。

方法

对 20 例肝移植受者进行了 39 次肝多普勒 US 检查。每次检查均以随机顺序进行 UltraFast 和常规肝多普勒 US。我们比较了肝血管的速度和时相、每次多普勒研究的持续时间、技术故障的发生情况以及临床决策的差异。

结果

肝血管的速度和阻力指数值在两种多普勒研究之间呈强正相关(平均 R=0.806;范围,0.710 至 0.924)。两种多普勒 US 检查中肝血管的时相相同。关于多普勒 US 检查的持续时间,UltraFast(251±99 秒)和常规(231±117 秒)多普勒研究之间无显著差异(P=0.306)。在 5 名呼吸不佳的受检者中,常规多普勒 US 的持续时间较长,UltraFast 多普勒 US(272±157 秒)所需时间短于常规多普勒 US(381±133 秒;P=0.005)。两种技术在技术故障和临床决策方面没有差异。

结论

UltraFast 多普勒 US 在肝移植后评估中对呼吸不佳的受检者具有优势,与常规多普勒 US 具有临床等效性。