Aldosterone determination: comparison of a RIA assay and a CLIA assay.

作者信息

Belaidi Nassima, Georges Agnès, Brossaud Julie, Corcuff Jean-Benoît

机构信息

Dpt of Nuclear Medicine, CHU de Bordeaux, F-33604 Pessac, France.

Dpt of Nuclear Medicine, CHU de Bordeaux, F-33604 Pessac, France; Univ. Bordeaux, Nutrition et Neurobiologie intégrée, UMR 1286, F-33000 Bordeaux, France.

出版信息

Clin Biochem. 2015 Jan;48(1-2):89-92. doi: 10.1016/j.clinbiochem.2014.10.008. Epub 2014 Oct 30.

Abstract

OBJECTIVE

To extend knowledge about the clinical performances of a new chemiluminescent immunoassay (CLIA) for aldosterone set up in available analysers.

DESIGN AND PATIENTS

We compared the results of a RIA assay to those of a CLIA assay in 198 serum and 80 urine samples from patients in endocrine and hypertension departments. Furthermore, for serum samples the concordance of results for postural tests was analysed.

RESULTS

RIA and CLIA aldosterone serum concentration was linearly correlated with a slope of 0.988 and an intercept of 70.4pmol/L. The variations of aldosterone serum concentration obtained with the two assays during postural tests were very consistent. There was no significant difference of aldosterone concentrations after thawing with the CLIA assay. RIA and CLIA aldosterone urine concentrations were linearly correlated with a slope of 0.787 and an intercept of -2.64nmol/L. Omitting the preservative boric acid from urine samples did not modify aldosterone concentration at least up to 48h after collection.

CONCLUSION

The RIA and CLIA assays were well correlated for the most useful serum samples. It is well suited to circumvent isotopic assays with the throughput of available analysers.

摘要

目的

拓展对在现有分析仪上建立的新型醛固酮化学发光免疫分析(CLIA)临床性能的认识。

设计与患者

我们比较了放射免疫分析(RIA)法与CLIA法对内分泌科和高血压科患者的198份血清样本及80份尿液样本的检测结果。此外,针对血清样本分析了体位试验结果的一致性。

结果

RIA法和CLIA法检测的醛固酮血清浓度呈线性相关,斜率为0.988,截距为70.4pmol/L。在体位试验期间,两种检测方法测得的醛固酮血清浓度变化非常一致。CLIA法检测解冻后的醛固酮浓度无显著差异。RIA法和CLIA法检测的醛固酮尿液浓度呈线性相关,斜率为0.787,截距为 -2.64nmol/L。尿液样本中省略防腐剂硼酸至少在采集后48小时内不会改变醛固酮浓度。

结论

对于大多数有用的血清样本,RIA法和CLIA法检测结果相关性良好。它非常适合利用现有分析仪的通量来规避同位素检测。