Anal Chim Acta. 2015 Jan 1;853:342-350. doi: 10.1016/j.aca.2014.10.041. Epub 2014 Oct 30.
A novel 8-electrode array as stir bar was designed for selective extraction of trace level exogenous estrogens from food samples, followed by liquid desorption and HPLC-photodiode array detection. The array consisted of 8 screen-printed electrodes and each electrode was modified with Fe3O4@meso-/macroporous TiO2 microspheres and molecularly imprinted film (m-TiMIF). The fabrication of the imprinted film coating was very simple without organic solvents and chemical grafting. Both bisphenol A (BPA) and diethylstilbestrol (DES) were employed as templates in m-TiMIF fabrication in order to enrich both targets simultaneously. Interestingly, the imprinted stir bar array showed higher extraction capacity and selectivity for BPA and DES than the non-imprinted counterpart. Meanwhile, it exhibited fast adsorption and desorption kinetics due to increased mass transport in the ultra-thin film. Importantly, the m-TiMIF coating was robust enough for at least 20 uses without obvious alteration in extraction performance. The main parameters affecting the extraction efficiency, including stir speeding, sample pH, ionic strength, extraction time, desorption solvent and time, were optimized. Under optimal experimental conditions, the limits of detection (S/N=3) of the developed method were 0.28 and 0.47 μg L(-1) for BPA and DES respectively, with enrichment factors of 32.6 and 52.8-fold. The linear ranges were 3.0-1500 μg L(-1) and 4.0-1500 μg L(-1) for BPA and DES, respectively. The m-TiMIF-coating conferred better recovery and selectivity, compared with the commercial stir bar coating. The new method was successfully applied to assess BPA and DES in pork and chicken samples with satisfactory recovery.
一种新型 8 电极搅拌棒作为搅拌棒,用于从食品样品中选择性提取痕量外源性雌激素,然后进行液相解吸和高效液相色谱-光电二极管阵列检测。该阵列由 8 个丝网印刷电极组成,每个电极都用 Fe3O4@介孔/大孔 TiO2 微球和分子印迹膜(m-TiMIF)修饰。印迹膜涂层的制备非常简单,无需有机溶剂和化学接枝。双酚 A(BPA)和己烯雌酚(DES)都被用作 m-TiMIF 制备中的模板,以便同时富集两种目标物。有趣的是,印迹搅拌棒阵列对 BPA 和 DES 的萃取容量和选择性均高于非印迹对照物。同时,由于超薄膜中增加了质量传递,它表现出更快的吸附和解吸动力学。重要的是,m-TiMIF 涂层足够坚固,可至少使用 20 次,而萃取性能无明显变化。优化了影响萃取效率的主要参数,包括搅拌速度、样品 pH 值、离子强度、萃取时间、解吸溶剂和时间。在最佳实验条件下,该方法对 BPA 和 DES 的检出限(S/N=3)分别为 0.28 和 0.47μg/L,富集因子分别为 32.6 和 52.8 倍。线性范围分别为 3.0-1500μg/L 和 4.0-1500μg/L。与商业搅拌棒涂层相比,m-TiMIF 涂层具有更好的回收率和选择性。该新方法成功应用于评估猪肉和鸡肉样品中的 BPA 和 DES,回收率令人满意。