Comparison of primary breast cancer size by mammography and sonography.

作者信息

Wang Jian-Tao, Chang Li-Ming, Song Xin, Zhao Li-Xin, Li Jun-Tao, Zhang Wei-Guo, Ji Ying-Bin, Cai Li-Na, Di Wei, Yang Xin-Yu

机构信息

X-Ray Department, Tangshan Gongren Hospital, Tangshan, Hebei, China E-mail :

出版信息

Asian Pac J Cancer Prev. 2014;15(22):9759-61. doi: 10.7314/apjcp.2014.15.22.9759.

Abstract

PURPOSE

To compare tumor size by mammography and sonography and align with pathological results in primary breast cancer cases.

MATERIALS AND METHODS

We retrospectively reviewed 95 primary breast cancer patients who underwent mammography and sonography from January 2011 to June 2012. The largest tumor diameter was chosen as sizing reference for each imaging modality. The measurements of mammography and sonography were considered concordant if they were within the measurement of pathological results±0.5 cm. Pearson's correlation coefficient was calculated for imaging results.

RESULTS

The range of the maximum diameter was 0.6 cm-10.5 cm and mean value was 3.81±2.04 cm by pathological results, 0.7 cm-12.4 cm and 3.99±2.19 cm by mammography, and 0.9 cm-11.0 cm and 3.63±2.01 cm by sonography, respectively. Sonography (R: 0.754), underestimated tumor size, but had a better correlation with pathological tumor size compared to mammography (R: 0.676), which overestimated tumor size.

CONCLUSIONS

Sonography is superior to mammography in assessment of primary breast cancer.