Hussein Lobna A, Magdy N, Abbas Mahmoud M
Pharmaceutical Analytical Chemistry Department, Faculty of Pharmacy, Ain Shams University, Abbassia, 11566, Cairo, Egypt.
Pharmaceutical Analytical Chemistry Department, Faculty of Pharmacy, Ain Shams University, Abbassia, 11566, Cairo, Egypt.
Spectrochim Acta A Mol Biomol Spectrosc. 2015 Mar 5;138:395-405. doi: 10.1016/j.saa.2014.11.073. Epub 2014 Nov 29.
Five simple, specific, accurate and precise UV-spectrophotometric methods are adopted for the simultaneous determination of Amprolium hydrochloride (AMP) and Ethopabate (ETH), a binary mixture with overlapping spectra, without preliminary separation. The first method is first derivative of the ratio spectra ((1)DD) for determination of AMP and ETH at 234.7nm and 306.8nm respectively with mean percentage recoveries 99.76±0.907 and 100.29±0.842 respectively. The second method is the mean centering of the ratio spectra for determination of AMP and ETH at 238.8nm and 313nm respectively with mean percentage recoveries 100.26±1.018 and 99.94±1.286 respectively. The third method is based on dual wavelength selection for determination of AMP and ETH at 235.3nm & 308nm and 244nm & 268.4nm respectively with mean percentage recoveries 99.30±1.097 and 100.03±1.065 respectively. The fourth method is ratio difference method for determination of AMP and ETH at 239nm & 310nm and 239nm & 313nm respectively with mean percentage recoveries 99.27±0.892 and 100.40±1.814 respectively. The fifth one is area under the curve (AUC) method where the areas between 235.6-243nm and 268.3-275nm are selected for determination of AMP and ETH with mean percentage recoveries 100.35±1.031 and 100.39±0.956 respectively. These methods are tested by analyzing synthetic mixtures of the two drugs and they are applied to their pharmaceutical veterinary preparation. Methods are validated according to the ICH guidelines and accuracy, precision and repeatability are found to be within the acceptable limit.
采用了五种简单、特定、准确且精确的紫外分光光度法,用于同时测定盐酸氨丙啉(AMP)和乙氧酰胺苯甲酯(ETH),这是一种光谱重叠的二元混合物,无需预先分离。第一种方法是比率光谱的一阶导数((1)DD),分别在234.7nm和306.8nm处测定AMP和ETH,平均回收率分别为99.76±0.907和100.29±0.842。第二种方法是比率光谱的平均中心化,分别在238.8nm和313nm处测定AMP和ETH,平均回收率分别为100.26±1.018和99.94±1.286。第三种方法基于双波长选择,分别在235.3nm和308nm以及244nm和268.4nm处测定AMP和ETH,平均回收率分别为99.30±1.097和100.03±1.065。第四种方法是比率差法,分别在239nm和310nm以及239nm和313nm处测定AMP和ETH,平均回收率分别为99.27±0.892和100.40±1.814。第五种方法是曲线下面积(AUC)法,选择235.6 - 243nm和268.3 - 275nm之间的面积来测定AMP和ETH,平均回收率分别为100.35±1.031和100.39±0.956。通过分析两种药物的合成混合物对这些方法进行了测试,并将其应用于它们的兽用药物制剂中。根据国际人用药品注册技术协调会(ICH)指南对方法进行了验证,发现准确度、精密度和重复性均在可接受范围内。