Suppr超能文献

经桡动脉途径经皮治疗复杂冠状动脉病变用导引导管延长系统的有效性和安全性。

Usefulness and safety of a guide catheter extension system for the percutaneous treatment of complex coronary lesions by a transradial approach.

机构信息

Department of Cardiology, Hospital Clínico Universitario de Valencia, Valencia, Spain.

出版信息

Med Princ Pract. 2015;24(2):171-7. doi: 10.1159/000369620. Epub 2014 Dec 20.

Abstract

OBJECTIVE

The aim of this study was to describe our initial experience with the GuideLiner® catheter (Vascular Solutions Inc.) in the transradial treatment of complex lesions.

MATERIALS AND METHODS

The clinical, angiographic and procedural data of percutaneous coronary interventions where GuideLiner was used during 2013 were collected. The transradial approach was used in all cases. The indication for its use, efficacy and periprocedural complications were determined. Sixteen consecutive procedures (in 15 patients; 12 males and 3 females) were evaluated. The indication for the use of GuideLiner was a difficulty to advance and properly position a stent through a tortuous and/or calcified artery despite using high-support guide catheters or other useful techniques.

RESULTS

Of the 16 angiographic procedures, 14 (87.5%) were successful (stent deployment in 13 cases and a drug-eluting balloon in 1 case). Unsuccessful cases were a chronic total occlusion and a diffusely diseased left anterior descendant artery. A type B dissection of a proximal left circumflex artery was the only periprocedural complication.

CONCLUSION

Use of the GuideLiner was an effective and safe technique for the percutaneous treatment of complex coronary lesions in which the adequate progress of angioplasty devices had failed. GuideLiner was particularly helpful when using the transradial approach. Only one minor complication was recorded.

摘要

目的

本研究旨在描述我们在经桡动脉途径治疗复杂病变中使用 GuideLiner®导管(Vascular Solutions Inc.)的初步经验。

材料和方法

收集了 2013 年期间在经皮冠状动脉介入治疗中使用 GuideLiner 的临床、血管造影和程序数据。所有病例均采用经桡动脉途径。确定了其使用的适应证、疗效和围手术期并发症。评估了 16 例连续手术(15 例患者,12 例男性,3 例女性)。使用 GuideLiner 的适应证是尽管使用高支撑导引导管或其他有用技术,但仍难以通过迂曲和/或钙化的动脉推进和正确放置支架。

结果

在 16 例血管造影程序中,14 例(87.5%)成功(13 例支架置入,1 例药物洗脱球囊)。不成功的病例是慢性完全闭塞和弥漫性病变的左前降支。近端左回旋支的 B 型夹层是唯一的围手术期并发症。

结论

在经皮治疗因血管成形术装置推进失败而导致的复杂冠状动脉病变时,使用 GuideLiner 是一种有效且安全的技术。在经桡动脉途径中,GuideLiner 特别有帮助。仅记录到 1 例轻微并发症。