Torque loss under mechanical cycling of long-span zirconia and titanium-cemented and screw-retained implant-supported CAD/CAM frameworks.

作者信息

Gomes Érica Alves, Tiossi Rodrigo, Faria Adriana Cláudia Lapria, Rodrigues Renata C S, Ribeiro Ricardo Faria

出版信息

Clin Oral Implants Res. 2014 Dec;25(12):1395-402. doi: 10.1111/clr.12286.

Abstract

OBJECTIVES

This study evaluated the screw joint stability after cyclic loading of implant-supported titanium and zirconia CAD/CAM frameworks for fixed dental prostheses (FDPs) with different retention methods.

MATERIAL AND METHODS

Twenty four one-piece frameworks supported by six threaded implants placed in the maxilla were fabricated using a CAD/CAM technique (NeoShape). Dry-pressed porcelain crowns were luted to the frameworks to standardize the specimens. The specimens were then divided into four groups (n = 6) according to framework material (titanium or zirconia) and retention method for the prosthesis (cement- or screw-retained): G1, Ti-cemented; G2, Ti-screw-retained; G3, Zr-cemented; and G4, Zr-screw-retained. A digital torque ratchet was used to assess the initial preload removal torque. Torque was then reapplied and the specimens were submitted to a 200 N cyclic load, at a frequency of 2 Hz, underwater in controlled temperature of 37°, and for 1 × 106 cycles. An opposing lower dental arch was fabricated using bis-acrylic resin to simulate occlusal contacts in centric. After cyclic loading, postload removal torque was measured. Preload and postload torque loss was expressed as a percentage of the initial load. Data were submitted to a linear mixed-effects model for statistical significance (α = 0.05) to evaluate the effect of cyclic loading in the screw torque loss used with frameworks of different materials and retention methods.

RESULTS

Significant screw torque loss (%) was found for the tested groups (before/after cyclic loading, respectively): G1 (39.77/61.83), G2 (37.57/50.96), G3 (34.87/54.10), and G4 (47.56/73.50) (P < 0.05).

CONCLUSIONS

The screw removal torque was significantly reduced for all groups in this study after cyclic loading the specimens. Screw-retained zirconia specimens presented the highest torque loss before and after the cyclic loadings compared with the other specimens that were tested.

摘要

目的

本研究评估了采用不同固位方法的种植体支持的钛和氧化锆计算机辅助设计/计算机辅助制造(CAD/CAM)固定义齿(FDP)框架在循环加载后的螺钉连接稳定性。

材料与方法

使用CAD/CAM技术(NeoShape)制作24个由上颌中6个螺纹种植体支持的一体式框架。将干压瓷冠粘结到框架上以标准化标本。然后根据框架材料(钛或氧化锆)和义齿的固位方法(粘结或螺钉固位)将标本分为四组(n = 6):G1,钛粘结;G2,钛螺钉固位;G3,氧化锆粘结;G4,氧化锆螺钉固位。使用数字扭矩棘轮评估初始预紧力去除扭矩。然后重新施加扭矩,并将标本在37°的受控温度下于水下以2 Hz的频率施加200 N的循环载荷,持续1×106次循环。使用双丙烯酸树脂制作一个相对的下颌牙弓,以模拟正中咬合接触。循环加载后,测量加载后去除扭矩。预紧力和加载后扭矩损失以初始载荷的百分比表示。将数据提交至线性混合效应模型以进行统计学显著性评估(α = 0.05),以评估循环加载对不同材料和固位方法的框架所使用的螺钉扭矩损失的影响。

结果

测试组的螺钉扭矩损失(%)显著(循环加载前后分别为):G1(39.77/61.83),G2(37.57/50.96),G3(34.87/54.10)和G4(47.56/73.50)(P < 0.05)。

结论

在本研究中,对标本进行循环加载后,所有组的螺钉去除扭矩均显著降低。与其他测试标本相比,螺钉固位的氧化锆标本在循环加载前后的扭矩损失最高。