Monte Carlo efficiency transfer method was used to determine the full energy peak efficiency of a coaxial n-type HPGe detector. The efficiencies calibration curves for three Certificate Reference Materials were determined by efficiency transfer using a (152)Eu reference source. The efficiency values obtained after efficiency transfer were used to calculate the activity concentration of the radionuclides detected in the three materials, which were measured in a low-background gamma spectrometry system. Reported and calculated activity concentration show a good agreement with mean deviations of 5%, which is satisfactory for environmental samples measurement.
采用蒙特卡罗效率转移法测定同轴n型高纯锗探测器的全能峰效率。使用(152)Eu参考源通过效率转移确定了三种有证标准物质的效率校准曲线。效率转移后获得的效率值用于计算在低本底γ能谱系统中测量的三种物质中检测到的放射性核素的活度浓度。报告的和计算的活度浓度显示出良好的一致性,平均偏差为5%,这对于环境样品测量来说是令人满意的。