Spinczyk Dominik
Faculty of Biomedical Engineering, Silesian University of Technology, Zabrze, Poland.
Wideochir Inne Tech Maloinwazyjne. 2014 Dec;9(4):531-6. doi: 10.5114/wiitm.2014.45048. Epub 2014 Sep 5.
For percutaneous abdomen intervention (e.g. liver radiofrequency (RF) tumor ablation, liver biopsy), surgeons lack real-time visual feedback about the location of the needle on planning images, typically computed tomography (CT). One difficulty lies in tracking and synchronizing both the tool movement and the patient breathing motion.
To verify the correspondence between rigid registration fiducial registration error signal and breathing phase.
Designed markers that are clearly visible both in planning CT and on the patient during the intervention are proposed. Registration and breathing synchronization is then performed by a point-based approach. The method was tested in a clinical environment on 10 patients with liver cancer using 3D abdominal CT in the exhale position. Median rigid fiducial registration error (FRE) in the breathing cycle was used as a criterion to distinguish the inhale and exhale phase.
The correlation between breathing phase and FRE value can be observed for every patient. We obtained mean median FRE equal to 9.37 mm in exhale positions and 15.56 mm in the whole breathing cycle.
The presented real time approach, based on FRE calculation, was integrated in the clinical pipeline, and can help to select the best respiratory phase for needle insertion for percutaneous abdomen intervention, in cases where only 3D CT is performed. Moreover, this method allows semi-automated rigid registration to establish the correspondence between preoperative patient anatomical model and patient position.
对于经皮腹部介入手术(例如肝脏射频(RF)肿瘤消融、肝脏活检),外科医生在规划图像(通常是计算机断层扫描(CT))上缺乏关于针位置的实时视觉反馈。一个难点在于跟踪和同步工具移动以及患者呼吸运动。
验证刚性配准基准配准误差信号与呼吸相位之间的对应关系。
提出了在规划CT和介入过程中的患者身上都清晰可见的设计标记。然后通过基于点的方法进行配准和呼吸同步。该方法在临床环境中对10例肝癌患者使用呼气位的三维腹部CT进行了测试。呼吸周期中的中位刚性基准配准误差(FRE)用作区分吸气和呼气阶段的标准。
每位患者均可观察到呼吸相位与FRE值之间的相关性。我们在呼气位获得的平均中位FRE等于9.37毫米,在整个呼吸周期中为15.56毫米。
所提出的基于FRE计算的实时方法已整合到临床流程中,并且在仅进行三维CT的情况下,有助于为经皮腹部介入手术选择最佳的针插入呼吸相位。此外,该方法允许半自动刚性配准以建立术前患者解剖模型与患者位置之间的对应关系。