Suppr超能文献

健康受试者下肢组织氧合监测中近红外光谱区域血红蛋白氧饱和度值及肢体与手臂比率的体内测量。

In vivo measurements of regional hemoglobin oxygen saturation values and limb-to-arm ratios of near-infrared spectroscopy for tissue oxygenation monitoring of lower extremities in healthy subjects.

作者信息

Boezeman Reinout Pe, Kelder Johannes C, Waanders Frans Gj, Moll Frans L, de Vries Jean-Paul Pm

机构信息

Department of Vascular Surgery, St Antonius Hospital, Nieuwegein, the Netherlands.

Department of Research and Developments, St Antonius Hospital, Nieuwegein, the Netherlands.

出版信息

Med Devices (Auckl). 2014 Dec 29;8:31-6. doi: 10.2147/MDER.S73103. eCollection 2015.

Abstract

OBJECTIVE

Near-infrared spectroscopy (NIRS) is a noninvasive technique that allows monitoring of regional hemoglobin oxygen saturation (rSO2) values and might have a role in the diagnosis of peripheral arterial disease. We assessed the reproducibility and inter-subject variability of rSO2 values and rSO2 limb-to-arm ratios (LARs) in lower extremities of healthy subjects.

METHODS

The rSO2 values and rSO2 LARs were calculated in eight healthy subjects without peripheral arterial disease. The rSO2 values were measured at rest at six fixed spots at each lower limb and a reference spot at each upper arm. NIRS provided the rSO2 values without involvement of any other processing technique. After measurements were completed, rSO2 LARs were calculated by dividing the rSO2 value of a lower extremity spot by the rSO2 value of the arm. Measurements were performed twice on 1 day and repeated on 4 different days.

RESULTS

Mean coefficients of variation of measurements of rSO2 values and rSO2 LARs at the same spot in the same subject were respectively less than 6% and 8% for every measurement spot over time. Coefficients of variation of measurements at the same spot between different subjects were less than 15% and 19% for every measurement spot respectively.

CONCLUSION

NIRS is an easily applicable, noninvasive tool for measurement of tissue oxygenation of lower extremities in healthy subjects. The reproducibility of rSO2 values and rSO2 LARs at the same measurement spot in the same subject is good.

摘要

目的

近红外光谱技术(NIRS)是一种无创技术,可用于监测局部血红蛋白氧饱和度(rSO2)值,可能在外周动脉疾病的诊断中发挥作用。我们评估了健康受试者下肢rSO2值和rSO2肢体与手臂比值(LARs)的可重复性和受试者间变异性。

方法

对8名无外周动脉疾病的健康受试者计算rSO2值和rSO2 LARs。在每个下肢的6个固定点以及每个上臂的一个参考点处静息状态下测量rSO2值。NIRS直接提供rSO2值,无需任何其他处理技术。测量完成后,通过将下肢某点的rSO2值除以上臂的rSO2值来计算rSO2 LARs。在1天内进行2次测量,并在4个不同日期重复测量。

结果

同一受试者同一测量点的rSO2值和rSO2 LARs测量的平均变异系数随时间推移在每个测量点分别小于6%和8%。不同受试者同一测量点测量的变异系数在每个测量点分别小于15%和19%。

结论

NIRS是一种易于应用的无创工具,可用于测量健康受试者下肢的组织氧合情况。同一受试者同一测量点的rSO2值和rSO2 LARs的可重复性良好。