Goto Atsushi, Nishikawa Jun, Kiyotoki Shu, Nakamura Munetaka, Nishimura Junichi, Okamoto Takeshi, Ogihara Hiroyuki, Fujita Yusuke, Hamamoto Yoshihiko, Sakaida Isao
Yamaguchi University Graduate School of Medicine, Department of Gastroenterology and Hepatology, Minamikogushi 1-1-1, Ube, Yamaguchi 755-8505, Japan.
Yamaguchi University Graduate School of Medicine, Department of Biomolecular Engineering, Tokiwadai 2-16-1, Ube, Yamaguchi 755-8611, Japan.
J Biomed Opt. 2015 Jan;20(1):016017. doi: 10.1117/1.JBO.20.1.016017.
Hyperspectral imaging (HSI) is a new technology that obtains spectroscopic information and renders it in image form. This study examined the difference in the spectral reflectance (SR) of gastric tumors and normal mucosa recorded with a hyperspectral camera equipped with HSI technology and attempted to determine the specific wavelength that is useful for the diagnosis of gastric cancer. A total of 104 gastric tumors removed by endoscopic submucosal dissection from 96 patients at Yamaguchi University Hospital were recorded using a hyperspectral camera. We determined the optimal wavelength and the cut-off value for differentiating tumors from normal mucosa to establish a diagnostic algorithm. We also attempted to highlight tumors by image processing using the hyperspectral camera's analysis software. A wavelength of 770 nm and a cut-off value of 1/4 the corrected SR were selected as the respective optimal wavelength and cut-off values. The rates of sensitivity, specificity, and accuracy of the algorithm's diagnostic capability were 71%, 98%, and 85%, respectively. It was possible to enhance tumors by image processing at the 770-nm wavelength. HSI can be used to measure the SR in gastric tumors and to differentiate between tumorous and normal mucosa.
高光谱成像(HSI)是一种获取光谱信息并将其以图像形式呈现的新技术。本研究检测了配备HSI技术的高光谱相机记录的胃肿瘤与正常黏膜的光谱反射率(SR)差异,并试图确定对胃癌诊断有用的特定波长。使用高光谱相机记录了山口大学医院96例患者经内镜黏膜下剥离术切除的104个胃肿瘤。我们确定了区分肿瘤与正常黏膜的最佳波长和临界值,以建立诊断算法。我们还尝试使用高光谱相机的分析软件通过图像处理突出显示肿瘤。分别选择770nm波长和校正后SR的1/4作为最佳波长和临界值。该算法诊断能力的敏感性、特异性和准确性分别为71%、98%和85%。在770nm波长下通过图像处理可以增强肿瘤显示。HSI可用于测量胃肿瘤的SR,并区分肿瘤性和正常黏膜。