Suppr超能文献

门控心血池单光子发射计算机断层扫描评估滤过对右心室功能的影响

Effects of filtration on right ventricular function by the gated blood pool SPECT.

作者信息

Mohseni Samane, Kamali-Asl Alireza, Bitarafan-Rajabi Ahmad, Entezarmahdi Seyed Mohammad, Shahpouri Zohre, Yaghoobi Nahid

机构信息

Radiation Medicine Engineering Department, Shahid Beheshti University, GC, Velenjak, Tehran, Iran.

出版信息

Ann Nucl Med. 2015 May;29(4):384-90. doi: 10.1007/s12149-015-0949-7. Epub 2015 Jan 31.

Abstract

OBJECTIVE

Gated blood po ol single photon emission computed tomography (GBPS) offers the possibility of obtaining additional functional information from blood pool studies, including evaluation of left and right ventricular function simultaneously. The calculation of ventricular volumes based on the identification of the endocardial surface would be influenced by the spatial resolution in the reconstructed images. This study was performed to evaluate the effect of different filters on the right ventricular function.

METHODS

The normal four-dimensional (4-D) NURBS-based cardiac-torso (NCAT) phantom with known right ventricular volume and ejection fraction was generated. The SIMIND Monte Carlo program was used to create projections. The studies were reconstructed by FBP and post-processing filtration such as Butterworth, Hanning, Shepp-Logan, Metz and Wiener in different statuses (cutoff and order). Using the Cedars-Sinai QBS (quantitative blood pool SPECT) package, the ventricular functional parameters were computed. The calculated values were analyzed and compared with the normal NCAT results.

RESULTS

The results implied that the calculated right ventricular end diastolic volume (RVEDV) by Butterworth filtration (cutoff frequency = 0.3) agreed more with the NCAT Phantom characteristics [relative difference percentage (RDP) = 1.2 %], while the maximum accordance in the calculation of the RV ejection fraction (EF) (RDP = 3 %) was observed by Metz filter (FWHM 20 pixel). Also, the results of this study demonstrate that the Butterworth filter provided the most stable values (cutoff frequency = 0.4-0.5) in the estimation of RVEDV (RDP = 7.5 %). The Hanning and Shepp-Logan filters produced a much larger RDP, particularly in low frequency (41.1 and 21.5 %, respectively) compared to other filters.

CONCLUSIONS

This study demonstrated that the operation of different filters has a severe effect in computing right ventricular volume. The resolution recovery and Butterworth filters tend to give more comparable ventricular volumes with the actual normal NCAT value. Further evaluation using a large clinical database is underway to evaluate the optimum protocol in a clinical setting.

摘要

目的

门控血池单光子发射计算机断层扫描(GBPS)提供了从血池研究中获取额外功能信息的可能性,包括同时评估左心室和右心室功能。基于心内膜表面识别来计算心室容积会受到重建图像空间分辨率的影响。本研究旨在评估不同滤波器对右心室功能的影响。

方法

生成具有已知右心室容积和射血分数的正常基于非均匀有理B样条(NURBS)的四维(4-D)心脏躯干(NCAT)体模。使用SIMIND蒙特卡罗程序创建投影。通过傅里叶反投影(FBP)和诸如巴特沃斯、汉宁、谢泼德-洛根、梅茨和维纳等后处理滤波在不同状态(截止频率和阶数)下对研究进行重建。使用雪松西奈定量血池SPECT(QBS)软件包计算心室功能参数。分析计算值并与正常NCAT结果进行比较。

结果

结果表明,巴特沃斯滤波(截止频率 = 0.3)计算的右心室舒张末期容积(RVEDV)与NCAT体模特征更相符[相对差异百分比(RDP)= 1.2%],而梅茨滤波器(半高宽20像素)在计算右心室射血分数(EF)时观察到最大一致性(RDP = 3%)。此外,本研究结果表明,巴特沃斯滤波器(截止频率 = 0.4 - 0.5)在估计RVEDV时提供了最稳定的值(RDP = 7.5%)。与其他滤波器相比,汉宁和谢泼德-洛根滤波器产生的RDP要大得多,尤其是在低频时(分别为41.1%和21.5%)。

结论

本研究表明,不同滤波器的操作在计算右心室容积方面有严重影响。分辨率恢复滤波器和巴特沃斯滤波器往往能给出与实际正常NCAT值更具可比性的心室容积。正在使用大型临床数据库进行进一步评估,以评估临床环境中的最佳方案。