Wanjian Gu, Liang Ge, Mingde Ji, Qing Yu, Chunbing Zhang
Clin Lab. 2014;60(10):1627-34. doi: 10.7754/clin.lab.2014.131019.
The aim of the study was to perform an internal comparison of blood cell analyses: white blood cell (WBC) and platelet (PLT) counts in a clinical laboratory.
Instrument comparison: Based on EP9-A2 of the Clinical Laboratory and Standard Institute (CLSI) and the statistics program of External Quality Assessment (EQA), we set 1/2CLIA'88 as our quality standard and used a Beckman coulter LH750 as the reference. We then compared this to another 4 blood analyzers: the Beckman Coulter Gen-S, ABX pentra 60, and the Sysmex XT2000i and XS-800i, including the Beckman Coulter Gen-S in manual mode. Personnel comparison: WBC and PLT counts were performed on five samples by 13 technicians at random. Two statistics programs were used to calculate the bias of each technician. The results were compared with the 1/2 CLIA'88 standard to evaluate staff performance.
All instrument comparisons met our quality standards of WBC < 7.5% and PLT < 12.5%. However, five technicians did not achieve satisfactory results of the WBC count in the first assessment. The coefficient of variation (CV) among qualified staff assessing five samples ranged from 8.51% to 12.2%, but all exceeded the 1/2 CLIA'88 quality standard; however, the CVs of 2 samples among qualified staff were approximately 20%, but after strict training this variability reduced significantly to approximately 5%.
Instrument comparisons demonstrated achievement of quality goals under rigorous calibration and quality control, but CVs were comparatively large. This improved following strict training, so a laboratory should emphasize the training and assessment of staff and evaluation of the actual situation when selecting a qualification program.
本研究旨在对临床实验室中的血细胞分析进行内部比较:白细胞(WBC)计数和血小板(PLT)计数。
仪器比较:基于临床实验室标准化协会(CLSI)的EP9-A2和外部质量评估(EQA)的统计程序,我们将1/2 CLIA'88作为质量标准,并使用贝克曼库尔特LH750作为参考。然后将其与另外4台血液分析仪进行比较:贝克曼库尔特Gen-S、ABX pentra 60以及Sysmex XT2000i和XS-800i,其中包括手动模式下的贝克曼库尔特Gen-S。人员比较:13名技术人员对5个样本进行白细胞和血小板计数。使用两个统计程序计算每个技术人员的偏差。将结果与1/2 CLIA'88标准进行比较以评估工作人员的表现。
所有仪器比较均符合我们白细胞<7.5%和血小板<12.5%的质量标准。然而,在首次评估中,有5名技术人员的白细胞计数未达到满意结果。合格人员对5个样本进行评估时的变异系数(CV)范围为8.51%至12.2%,但均超过了1/2 CLIA'88质量标准;不过,合格人员对2个样本的CV约为20%,但经过严格培训后,这种变异性显著降低至约5%。
仪器比较表明在严格校准和质量控制下实现了质量目标,但CV相对较大。经过严格培训后有所改善,因此实验室在选择资格认证程序时应强调对工作人员的培训和评估以及对实际情况的评估。