使用压力传感器测量高强度聚焦超声换能器的声功率

Acoustic power measurement of high-intensity focused ultrasound transducer using a pressure sensor.

作者信息

Zhou Yufeng

机构信息

Division of Engineering Mechanics, School of Mechanical and Aerospace Engineering, Nanyang Technological University, Singapore 639798, Singapore .

出版信息

Med Eng Phys. 2015 Mar;37(3):335-40. doi: 10.1016/j.medengphy.2015.01.014. Epub 2015 Feb 7.

Abstract

The acoustic power of high-intensity focused ultrasound (HIFU) is an important parameter that should be measured prior to each treatment to guarantee effective and safe outcomes. A new calibration technique was developed that involves estimating the pressure distribution, calculating the acoustic power using an underwater pressure blast sensor, and compensating the contribution of harmonics to the acoustic power. The output of a clinical extracorporeal HIFU system (center frequency of ~1 MHz, p+ = 2.5-57.2 MPa, p(-) = -1.8 to -13.9 MPa, I(SPPA) = 513-22,940 W/cm(2), -6 dB size of 1.6 × 10 mm: lateral × axial) was measured using this approach and then compared with that obtained using a radiation force balance. Similarities were found between each method at acoustic power ranging from 18.2 W to 912 W with an electrical-to-acoustic conversion efficiency of ~42%. The proposed method has advantages of low weight, smaller size, high sensitivity, quick response, high signal-to-noise ratio (especially at low power output), robust performance, and easy operation of HIFU exposimetry measurement.

摘要

高强度聚焦超声(HIFU)的声功率是一个重要参数,每次治疗前都应进行测量,以确保治疗效果有效且安全。开发了一种新的校准技术,该技术涉及估计压力分布、使用水下压力冲击传感器计算声功率以及补偿谐波对声功率的贡献。使用这种方法测量了临床体外HIFU系统的输出(中心频率约为1 MHz,p+ = 2.5 - 57.2 MPa,p(-) = -1.8至-13.9 MPa,I(SPPA) = 513 - 22,940 W/cm²,-6 dB尺寸为1.6×10 mm:横向×轴向),然后将其与使用辐射力平衡获得的结果进行比较。在声功率范围为18.2 W至912 W、电声转换效率约为42%时,两种方法之间发现了相似之处。所提出的方法具有重量轻、尺寸小、灵敏度高、响应快、信噪比高(特别是在低功率输出时)、性能稳健以及HIFU剂量测量操作简便等优点。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验