Huang Jie, Lu Youli, Wan Qian, Zhang Meiwei, Pei Qi, Zhang Mengqi, Liu Gangyi, Yang Guoping
Center of Clinical Pharmacology, Third Xiangya Hospital, Central South University, Changsha, Hunan 410013, China.
Drug Clinical Research Centre of Shanghai Xuhui Central Hospital, Shanghai 200031, China.
J Chromatogr Sci. 2015 Sep;53(8):1303-9. doi: 10.1093/chromsci/bmv009. Epub 2015 Feb 16.
A highly selective and sensitive liquid chromatography-tandem mass spectrometry method was developed and validated for the simultaneous determination of trantinterol and one of its major metabolites, 1-carbonyl trantinterol, in human plasma. An Oasis MCX 96-well solid-phase extraction cartridge and a SeQuantTM ZIC(®)-HILIC LC column were used for sample preparation and chromatographic separation, respectively. The analytes were monitored by a QTrap 5500 mass spectrometer with positive electrospray ionization. Multiple reaction monitoring was used for quantification using the precursor to product ion pairs of m/z 311.1 → 237.9 (trantinterol), m/z 325.1 → 251.9 (1-carbonyl trantinterol) and m/z 368.4 → 294.0 (bambuterol as internal standard). The assay had a calibration range from 0.2 to 50 pg/mL and a lower limit of quantification of 0.2 pg/mL for both trantinterol and 1-carbonyl trantinterol. The inter-day and intra-day precisions were <12.0% and the accuracies were within the range of 87.1-111%. The mean recovery ranged from 82.0 to 97.7% and internal standard normalized matrix effect from 0.813 to 0.899. The analytes were stable under all tested conditions. This validated method was successfully applied to a pilot pharmacokinetic study in healthy subjects administered a single 50 μg oral dose.