Howe Andrew J, McKeag Nicholas A, Wilson Carol M, Ashfield Kyle P, Roberts Michael J
Department of Cardiology, Royal Victoria Hospital, Belfast, UK.
J Cardiovasc Electrophysiol. 2015 Jun;26(6):686-9. doi: 10.1111/jce.12638. Epub 2015 Apr 18.
Implantable cardioverter defibrillator (ICD) lead insulation failure and conductor externalization have been increasingly reported. The 7.8F silicon-insulated Linox SD and Linox S ICD leads (Biotronik, Berlin, Germany) were released in 2006 and 2007, respectively, with an estimated 85,000 implantations worldwide. A 39-year-old female suffered an out-of-hospital ventricular fibrillation (VF) arrest with successful resuscitation. An ICD was implanted utilizing a single coil active fixation Linox(Smart) S lead (Biotronik, Berlin, Germany). A device-triggered alert approximately 3 years after implantation confirmed nonphysiological high rate sensing leading to VF detection. A chest X-ray showed an abnormality of the ICD lead and fluoroscopic screening confirmed conductor externalization proximal to the defibrillator coil. In view of the combined electrical and fluoroscopic abnormalities, urgent lead extraction and replacement were performed. A review of Linox (Biotronik) and Vigila (Sorin Group, Milan, Italy) lead implantations within our center (n = 98) identified 3 additional patients presenting with premature lead failure, 2 associated with nonphysiological sensed events and one associated with a significant decrease in lead impedance. All leads were subsequently removed and replaced. This case provides a striking example of insulation failure affecting the Linox ICD lead and, we believe, is the first to demonstrate conductor externalization manifesting both electrical and fluoroscopic abnormalities.
植入式心脏复律除颤器(ICD)导线绝缘故障和导线外露的报道日益增多。7.8F硅绝缘的Linox SD和Linox S ICD导线(德国柏林百多力公司)分别于2006年和2007年上市,全球估计植入了85,000例。一名39岁女性在院外发生心室颤动(VF)骤停,复苏成功。植入了一根单线圈主动固定Linox(智能)S导线(德国柏林百多力公司)的ICD。植入后约3年,设备触发警报,证实存在非生理性高心率感知,导致VF检测。胸部X线显示ICD导线异常,透视筛查证实除颤器线圈近端导线外露。鉴于电学和透视检查均有异常,紧急进行了导线拔除和更换。回顾我们中心植入的Linox(百多力公司)和Vigila(意大利米兰索林集团)导线(n = 98),发现另外3例患者出现导线过早故障,2例与非生理性感知事件有关,1例与导线阻抗显著降低有关。随后所有导线均被拔除并更换。本病例提供了一个影响Linox ICD导线的绝缘故障的显著例子,我们认为,这是首例同时表现出电学和透视异常的导线外露病例。