Cardiology and Angiology I, Heart Center, Freiburg University, Freiburg, Germany.
Department of Radiology, University Hospital, Freiburg, Germany.
Europace. 2015 Oct;17(10):1526-32. doi: 10.1093/europace/euu410. Epub 2015 Mar 4.
Lesion formation during catheter ablation crucially depends on catheter-tissue contact. We sought to evaluate the impact of anatomical characteristics of the left atrium (LA) and the pulmonary veins (PVs) on contact force (CF) measurements.
An anatomical map of the LA was obtained in 25 patients prior to catheter ablation of atrial fibrillation. Contact force (operator blinded) and local bipolar electrogram amplitudes (EGM) were measured in eight pre-defined segments around the PVs. After unblinding, points with low CF (≤5 g) were corrected to CF >5 g, and the distance between points was measured. In a pre-procedural computed tomography of the heart, LA volume as well as sizes and circumferences of the PV ostia were measured and correlated to CF measurements. Four hundred and twenty-six points in eight pre-defined LA locations were assessed. Low CF (<5 g) was found in 25.0% (43.5%) of points superior, 33.3% (66.7%) anterior, 32.1% (44.4%) inferior, and 15.5% (15.9%) posterior to the right (left) PVs. The mean distance after correction was 5.8 ± 3.4 mm. Local bipolar electrogram amplitudes between low- and high-CF points did not differ (1.21 ± 1.54 vs. 1.13 ± 1.3 mV, P = ns). The mean CF at the left PVs was significantly lower than at the right PVs (7.91 ± 3.74 vs. 13.95 ± 6.34 g, P < 0.001), with the lowest CF anterior to the left PVs (5.2 ± 3.6 g). Contact force measurements did not correlate to LA volume, size, and circumference of the PVs.
Contact force during LA mapping significantly differs according to the location within the LA. These differences are independent of LA volume and anatomy of the PV ostia.
导管消融过程中病灶的形成主要取决于导管与组织的接触。我们旨在评估左心房(LA)和肺静脉(PV)的解剖特征对接触力(CF)测量的影响。
在 25 例房颤导管消融前,获得了 LA 的解剖图谱。在 PV 周围的八个预定义部位测量接触力(术者盲法)和局部双极电图幅度(EGM)。在揭盲后,将 CF 较低(≤5 g)的点校正为 CF>5 g,并测量两点之间的距离。在心脏的术前 CT 中,测量了 LA 容积以及 PV 口的大小和周长,并与 CF 测量值相关。评估了 8 个预定义 LA 部位的 426 个点。右(左)PV 上方、前方、下方和后方的 CF 较低(<5 g)的点分别占 25.0%(66.7%)、33.3%(44.4%)、32.1%(44.4%)和 15.5%(15.9%)。校正后的平均距离为 5.8 ± 3.4mm。低 CF 点与高 CF 点之间的局部双极电图幅度无差异(1.21 ± 1.54 与 1.13 ± 1.3 mV,P = ns)。左 PV 的平均 CF 明显低于右 PV(7.91 ± 3.74 与 13.95 ± 6.34 g,P<0.001),左 PV 前方的 CF 最低(5.2 ± 3.6 g)。CF 测量值与 LA 容积、PV 口大小和周长无相关性。
LA 标测过程中的接触力根据 LA 内的位置显著不同。这些差异与 LA 容积和 PV 口解剖结构无关。