不同射束方向对使用单光子发射计算机断层扫描识别的功能肺组织所接受的调强放射治疗剂量的影响。

Impact of different beam directions on intensity-modulated radiation therapy dose delivered to functioning lung tissue identified using single-photon emission computed tomography.

作者信息

Tian Qin, Zhang Fucheng, Wang Yanming, Qu Weiqiang

机构信息

Department of Radiotherapy, The General Hospital of Jinan Military Command, Jinan, Shandong Province, China ; Qin Tian and Fucheng Zhang contributed equally to this work.

Department of Gastroenterology, The General Hospital of Jinan Military Command, Jinan, Shandong Province, China ; Qin Tian and Fucheng Zhang contributed equally to this work.

出版信息

Contemp Oncol (Pozn). 2014;18(6):436-41. doi: 10.5114/wo.2014.46237. Epub 2014 Nov 12.

Abstract

AIM OF THE STUDY

To use different beam arrangements and numbers to plan intensity-modulated radiation therapy (IMRT) and investigate their effects on low and high radiation doses delivered to the functional lung, in order to reduce radiation-induced lung damage.

MATERIAL AND METHODS

Ten patients with stage I-III non-small cell lung carcinoma (NSCLC) underwent IMRT. Beam arrangements were selected on the basis of orientation and dose-volume histograms to create SPECT-guided IMRT plans that spared the functional lung and maintained target coverage. Four different plans, including CT-7, SPECT-7, SPECT-4, SPECT-5 with different beam arrangements, were used. The differences of conformity index (CI), heterogeneity index (HI) between the plans were analyzed, by using a paired t-test.

RESULTS

The seven-beam SPECT (SPECT-7) plan reduced the volume of the functional lung irradiated with at least 20 Gy (FV20) and 30 Gy (FV30) by 26.02% ±15.45% and 14.41% ±16.66%, respectively, as compared to the seven-beam computed tomography (CT-7) plan. The CI significantly differed between the SPECT-7 and SPECT-4 plans and between the SPECT-5 and SPECT-4 plans, but not between the SPECT-5 and SPECT-7 plans. The CIs in the SPECT-5 and SPECT-7 plans were better than that in the SPECT-4 plan. The heterogeneity index significantly differed among the three SPECT plans and was best in the SPECT-7 plan.

CONCLUSIONS

The incorporation of SPECT images into IMRT planning for NSCLC greatly affected beam angles and number of beams. Fewer beams and modified beam angles achieved similar or better IMRT quality. The low-dose volumes were lower in SPECT-4.

摘要

研究目的

采用不同的射束排列和数量来计划调强放射治疗(IMRT),并研究其对输送到功能性肺组织的低、高辐射剂量的影响,以减少放射性肺损伤。

材料与方法

10例I - III期非小细胞肺癌(NSCLC)患者接受IMRT。根据方向和剂量体积直方图选择射束排列,以创建SPECT引导的IMRT计划,该计划可保护功能性肺组织并维持靶区覆盖。使用了四种不同的计划,包括具有不同射束排列的CT - 7、SPECT - 7、SPECT - 4、SPECT - 5。采用配对t检验分析各计划之间的适形指数(CI)、异质性指数(HI)差异。

结果

与七射束计算机断层扫描(CT - 7)计划相比,七射束SPECT(SPECT - 7)计划使接受至少20 Gy(FV20)和30 Gy(FV30)照射的功能性肺组织体积分别减少了26.02%±15.45%和14.41%±16.66%。SPECT - 7与SPECT - 4计划之间以及SPECT - 5与SPECT - 4计划之间的CI有显著差异,但SPECT - 5与SPECT - 7计划之间无显著差异。SPECT - 5和SPECT - 7计划中的CI优于SPECT - 4计划。三种SPECT计划之间的异质性指数有显著差异,且在SPECT - 7计划中最佳。

结论

将SPECT图像纳入NSCLC的IMRT计划中极大地影响了射束角度和射束数量。较少的射束和调整后的射束角度可实现相似或更好的IMRT质量。SPECT - 4中的低剂量体积更低。