Sutton R, Perrins E J, Duffin E
Pacing Clin Electrophysiol. 1985 Jan;8(1):6-16. doi: 10.1111/j.1540-8159.1985.tb05716.x.
A diagramming system has been developed to assist in interpreting dual chamber pacemaker electrocardiograms. The system is based on the ladder diagram principle but uses two separate ladders, one for the ECG allowing proper representation of atrial A-V and ventricular activation, and a second ladder diagram below it for pacemaker events. The system is applicable to all current single and dual chamber pacing modes and is easy to draw. The ladder diagram allows the logical sequence of timing events to be easily illustrated, and pacemaker malfunction becomes clearly recognizable as a conflict between the ECG events ladder and the pacemaker timing ladder. The system is easy to apply and may eventually prove suitable for automatic generation by computer utilizing sophisticated pacemaker telemetry.
已开发出一种图解系统,以协助解读双腔起搏器心电图。该系统基于梯形图原理,但使用两个独立的梯形图,一个用于心电图,以便正确表示心房、房室和心室激动,另一个在其下方用于起搏器事件。该系统适用于所有当前的单腔和双腔起搏模式,且易于绘制。梯形图能轻松展示计时事件的逻辑顺序,起搏器故障也能作为心电图事件梯形图与起搏器计时梯形图之间的冲突清晰识别出来。该系统易于应用,最终可能证明适用于利用复杂的起搏器遥测技术由计算机自动生成。