Precision, high dose radiotherapy: helium ion treatment of uveal melanoma.

作者信息

Saunders W M, Char D H, Quivey J M, Castro J R, Chen G T, Collier J M, Cartigny A, Blakely E A, Lyman J T, Zink S R

出版信息

Int J Radiat Oncol Biol Phys. 1985 Feb;11(2):227-33. doi: 10.1016/0360-3016(85)90143-9.

Abstract

We report on 75 patients with uveal melanoma who were treated by placing the Bragg peak of a helium ion beam over the tumor volume. The technique localizes the high dose region very tightly around the tumor volume. This allows critical structures, such as the optic disc and the macula, to be excluded from the high dose region as long as they are 3 to 4 mm away from the edge of the tumor. Careful attention to tumor localization, treatment planning, patient immobilization and treatment verification is required. With a mean follow-up of 22 months (3 to 60 months) we have had only five patients with a local recurrence, all of whom were salvaged with another treatment. Pretreatment visual acuity has generally been preserved as long as the tumor edge is at least 4 mm away from the macula and optic disc. The only serious complication to date has been an 18% incidence of neovascular glaucoma in the patients treated at our highest dose level. Clinical results and details of the technique are presented to illustrate potential clinical precision in administering high dose radiotherapy with charged particles such as helium ions or protons.