Song Manli, Chen Jyh-Horng, Chen Ji, Lin In-Tsang
State Key Laboratory of Molecular Vaccinology and Molecular Diagnostics, School of Public Health, Xiamen University, Xiamen Fujian, China; Center for Molecular Imaging and Translational Medicine, School of Public Health, Xiamen University, Xiamen Fujian, China.
Interdisciplinary MRI/MRS Lab, Department of Electrical Engineering, National Taiwan University, Taipei, Taiwan; Graduate Institute of Biomedical Electronics and Bioinformatics, National Taiwan University, Taipei, Taiwan.
PLoS One. 2015 Mar 26;10(3):e0118892. doi: 10.1371/journal.pone.0118892. eCollection 2015.
This study attempts to compare the signal-to-noise ratio (SNR) of the 40 mm High-Temperature Superconducting (HTS) surface resonator at 77 K and the 35 mm commercial quadrature (QD) surface resonator at 300 K in a 3 Tesla (T) MRI imager. To aquire images for the comparison, we implemented a phantom experiment using the 40 mm diameter Bi2Sr2Ca2Cu3Ox (Bi-2223) HTS surface resonator, the 35 mm commercial QD surface resonator and the 40 mm professionally-made copper surface resonator. The HTS surface resonator at 77 K provided a 1.43-fold SNR gain over the QD surface resonator at 300 K and provided a 3.84-fold SNR gain over the professionally-made copper surface resonator at 300 K on phantom images. The results agree with the predictions, and the difference between the predicted SNR gains and measured SNR gains is 1%. Although the geometry of the HTS surface resonator is different from the QD surface resonator, its SNR is still higher. The results demonstrate that a higher image quality can be obtained with the HTS surface resonator at 77 K. With the HTS surface resonator, the SNR can be improved, suggesting that the HTS surface resonator is a potentially helpful diagnostic tool for MRI imaging in various applications.
本研究试图在一台3特斯拉(T)的磁共振成像(MRI)仪中,比较77 K时40毫米高温超导(HTS)表面谐振器与300 K时35毫米商用正交(QD)表面谐振器的信噪比(SNR)。为获取用于比较的图像,我们使用直径40毫米的Bi2Sr2Ca2Cu3Ox(Bi-2223)HTS表面谐振器、35毫米商用QD表面谐振器和40毫米专业制作的铜表面谐振器进行了体模实验。在体模图像上,77 K的HTS表面谐振器比300 K的QD表面谐振器的信噪比提高了1.43倍,比300 K的专业制作的铜表面谐振器的信噪比提高了3.84倍。结果与预测相符,预测的信噪比增益与测量的信噪比增益之间的差异为1%。尽管HTS表面谐振器的几何形状与QD表面谐振器不同,但其信噪比仍然更高。结果表明,77 K的HTS表面谐振器可获得更高的图像质量。使用HTS表面谐振器可以提高信噪比,这表明HTS表面谐振器在各种应用中对MRI成像可能是一种有用的诊断工具。