Chhabra Lovely, Khalid Nauman, Kluger Jeffrey
Department of Cardiovascular Medicine, Hartford Hospital, University of Connecticut School of Medicine, Hartford, Connecticut.
Proc (Bayl Univ Med Cent). 2015 Apr;28(2):200-3. doi: 10.1080/08998280.2015.11929229.
Appropriate sensing is essential for the normal functioning of an implantable cardioverter defibrillator (ICD). T-wave oversensing is a potential clinical problem in ICD patients that may result in inappropriate shocks. Oversensing may have various underlying causes and can be treated with noninvasive or invasive means. We present the case of a 45-year-old man presenting with shock storm as a result of T-wave oversensing. Workup revealed a hemodynamically significant stenosis of the left anterior descending artery treated with percutaneous coronary intervention and drug-eluting stent placement. This resulted in the resolution of T-wave missensing and restoration of normal ICD functioning.
恰当的感知对于植入式心脏复律除颤器(ICD)的正常运行至关重要。T波感知过度是ICD患者中一个潜在的临床问题,可能导致不适当的电击。感知过度可能有多种潜在原因,可通过非侵入性或侵入性手段进行治疗。我们报告一例45岁男性因T波感知过度出现电击风暴的病例。检查发现左前降支存在具有血流动力学意义的狭窄,经皮冠状动脉介入治疗并植入药物洗脱支架。这使得T波误感知得到解决,ICD功能恢复正常。