Soliman K, Bakkari M
Department of Medical Physics, Prince Sultan Military Medical City, P.O. Box 7897, Riyadh 11159, Saudi Arabia
Department of Medical Physics, Prince Sultan Military Medical City, P.O. Box 7897, Riyadh 11159, Saudi Arabia.
Radiat Prot Dosimetry. 2015 Jul;165(1-4):373-5. doi: 10.1093/rpd/ncv126. Epub 2015 Apr 6.
Based on manufacturer specifications, radiochromic films are sensitive enough to be used for dosimetry in digital mammography (DM). The aim of this work was to study the feasibility of measuring entrance surface dose (ESD) distribution using Gafchromic XR-QA2 films. The films were irradiated following a standard clinical two-view screening mammography protocol using a full-field digital mammography (FFDM) imaging system. The films were then digitised using a flatbed scanner. The calibration curve relating the readings from a calibrated ionisation chamber and the films' net optical density (NOD) could not be obtained. The examination of the calibration data revealed non-sensitivity of the films to resolve dose differences below 20 mGy at 28 kVp. Therefore, radiochromic films were found not to be suitable for measuring ESD profiles in DM. A 2D map of the NOD of the irradiated films obtained using in-house developed MATLAB computer program is presented.