检测在传统数字乳腺摄影中使用放射变色胶片测量皮肤入口剂量分布的相关性。

Examination of the relevance of using radiochromic films in measuring entrance skin dose distribution in conventional digital mammography.

作者信息

Soliman K, Bakkari M

机构信息

Department of Medical Physics, Prince Sultan Military Medical City, P.O. Box 7897, Riyadh 11159, Saudi Arabia

Department of Medical Physics, Prince Sultan Military Medical City, P.O. Box 7897, Riyadh 11159, Saudi Arabia.

出版信息

Radiat Prot Dosimetry. 2015 Jul;165(1-4):373-5. doi: 10.1093/rpd/ncv126. Epub 2015 Apr 6.

Abstract

Based on manufacturer specifications, radiochromic films are sensitive enough to be used for dosimetry in digital mammography (DM). The aim of this work was to study the feasibility of measuring entrance surface dose (ESD) distribution using Gafchromic XR-QA2 films. The films were irradiated following a standard clinical two-view screening mammography protocol using a full-field digital mammography (FFDM) imaging system. The films were then digitised using a flatbed scanner. The calibration curve relating the readings from a calibrated ionisation chamber and the films' net optical density (NOD) could not be obtained. The examination of the calibration data revealed non-sensitivity of the films to resolve dose differences below 20 mGy at 28 kVp. Therefore, radiochromic films were found not to be suitable for measuring ESD profiles in DM. A 2D map of the NOD of the irradiated films obtained using in-house developed MATLAB computer program is presented.

摘要

根据制造商的规格,放射变色胶片足够灵敏,可用于数字乳腺摄影(DM)中的剂量测定。这项工作的目的是研究使用Gafchromic XR-QA2胶片测量入射表面剂量(ESD)分布的可行性。按照标准临床双视图筛查乳腺摄影方案,使用全场数字乳腺摄影(FFDM)成像系统对胶片进行照射。然后使用平板扫描仪对胶片进行数字化处理。无法获得将校准电离室的读数与胶片的净光密度(NOD)相关联的校准曲线。对校准数据的检查表明,在28 kVp时,胶片对分辨低于20 mGy的剂量差异不敏感。因此,发现放射变色胶片不适合测量DM中的ESD分布。展示了使用自行开发的MATLAB计算机程序获得的照射胶片的NOD二维图。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验