Bassey Bassey, Moreno Beatriz, Chapman Dean
Department of Physics & Engineering Physics, University of Saskatchewan, Saskatoon, Canada.
J Radiol Prot. 2015 Jun;35(2):383-90. doi: 10.1088/0952-4746/35/2/383. Epub 2015 Apr 23.
Protection and operational quantities as defined by the International Commission on Radiological Protection (ICRP) and the International Commission on Radiation Units and Measurements (ICRU) are the two sets of quantities recommended for use in radiological protection for external radiation. Since the '80s, the protection quantities have evolved from the concept of dose equivalent to effective dose equivalent to effective dose, and the associated conversion coefficients have undergone changes. In this work, the influence of three different versions of ICRP photon dose conversion coefficients in the synchrotron radiation shielding calculations of an experimental enclosure has been examined. The versions are effective dose equivalent (ICRP Publication 51), effective dose (ICRP Publication 74), and effective dose (ICRP Publication 116) conversion coefficients. The sources of the synchrotron radiation white beam into the enclosure were a bending magnet, an undulator and a wiggler. The ranges of photons energy from these sources were 10-200 keV for the bending magnet and undulator, and 10-500 keV for the wiggler. The design criterion aimed a radiation leakage less than 0.5 µSv h(-1) from the enclosure. As expected, larger conversion coefficients in ICRP Publication 51 lead to higher calculated dose rates. However, the percentage differences among the calculated dose rates get smaller once shielding is added, and the choice of conversion coefficients set did not affect the final shielding decision.
国际放射防护委员会(ICRP)和国际辐射单位与测量委员会(ICRU)定义的防护量和运算量是推荐用于外照射放射防护的两组量。自20世纪80年代以来,防护量已从剂量当量概念演变为有效剂量当量再到有效剂量,相关的转换系数也发生了变化。在这项工作中,研究了ICRP光子剂量转换系数的三个不同版本在一个实验屏蔽室同步辐射屏蔽计算中的影响。这三个版本分别是有效剂量当量(ICRP第51号出版物)、有效剂量(ICRP第74号出版物)和有效剂量(ICRP第116号出版物)转换系数。进入屏蔽室的同步辐射白光束源有一个弯曲磁铁、一个波荡器和一个摆动器。这些源的光子能量范围对于弯曲磁铁和波荡器是10 - 200 keV,对于摆动器是10 - 500 keV。设计标准旨在使屏蔽室的辐射泄漏小于0.5 µSv h⁻¹。正如预期的那样,ICRP第51号出版物中较大的转换系数导致计算出的剂量率更高。然而,一旦添加屏蔽,计算出的剂量率之间的百分比差异会变小,并且转换系数集的选择并不影响最终的屏蔽决策。