OBJECTIVES

To evaluate the efficiency of automatic respiratory gating (ARG) in reducing respiratory motion-induced artefacts from dynamic contrast-enhanced ultrasound (DCEUS) acquisitions and to assess the impact of ARG on DCEUS quantification parameters in patients with liver malignancies.

METHODS

Twenty-five patients with liver metastasis were imaged with DCEUS. The lognormal indicator dilution model was fitted on time-intensity curves extracted from hepatic lesions with and without the use of ARG and DCEUS quantification parameters were extracted. The goodness of fit was assessed using the coefficient of determination (R (2) LN ). The effect respiration had on the data was assessed using the respiration amplitude (RA) metric. Pearson's correlation coefficient (r) was used to assess the correlation between R (2) LN and RA with and without the use of ARG.

RESULTS

The RA parameter was strongly correlated with R (2) LN (r = -0.96, P = 7.412 × 10(-15)) and this correlation became weaker with ARG (r = -0.64, P = 5.449 × 10(-4)). ARG significantly influenced the values of the quantification parameters extracted (P ≤ 0.05). The RA was significantly decreased when ARG was used (P = 1.172 × 10(-6)).

CONCLUSIONS

ARG has a significant impact on the quantification parameters extracted and it has been shown to improve the accuracy of liver lesion DCEUS.

KEY POINTS

• ARG has a significant impact on DCEUS quantification parameters. • ARG can improve the modelling of liver lesion haemodynamics using DCEUS quantification. • ARG significantly reduces the respiration amplitude of DCEUS lesion time-intensity curves.

摘要

目的

评估自动呼吸门控(ARG)在减少动态对比增强超声(DCEUS)采集过程中呼吸运动引起的伪影的效率,并评估 ARG 对肝恶性肿瘤患者 DCEUS 定量参数的影响。

方法

对 25 例肝转移患者进行 DCEUS 成像。使用对数正态指示稀释模型拟合从肝病变中提取的时间-强度曲线,提取 DCEUS 定量参数。使用决定系数(R (2) LN )评估拟合优度。使用呼吸幅度(RA)指标评估呼吸对数据的影响。使用 Pearson 相关系数(r)评估 R (2) LN 和 RA 在使用和不使用 ARG 时的相关性。

结果

RA 参数与 R (2) LN 高度相关(r = -0.96,P = 7.412×10(-15)),并且这种相关性随着 ARG 的使用而减弱(r = -0.64,P = 5.449×10(-4))。ARG 显著影响提取的定量参数值(P≤0.05)。使用 ARG 时,RA 显著降低(P = 1.172×10(-6))。

结论

ARG 对提取的定量参数有显著影响,已证明它可以提高肝脏病变 DCEUS 的准确性。

关键点

• ARG 对 DCEUS 定量参数有显著影响。

• ARG 可以改善肝脏病变 DCEUS 血流动力学的建模。

• ARG 显著降低了 DCEUS 病变时间-强度曲线的呼吸幅度。