IEEE Trans Med Imaging. 2015 Oct;34(10):2186-90. doi: 10.1109/TMI.2015.2426684. Epub 2015 Apr 28.
Contrast agent enhanced cone beam computed tomography (CE-CBCT), a technique capable of high-resolution in vivo imaging with small radiation dose, has been applied successfully for clinical diagnostics of cartilage degeneration, i.e., osteoarthritis (OA). As an X-ray technique, CE-CBCT may also detect changes in mineral density of subchondral bone (volumetric bone mineral density, vBMD), known to be characteristic for OA. However, its feasibility for density measurements is not clear due to limited signal-to-noise ratio and contrast of CBCT images. In the present study, we created clinically applicable hydroxyapatite phantoms and determined vBMDs of cortical bone, trabecular bone, subchondral trabecular bone and subchondral plate of 10 cadaver (ex vivo) and 10 volunteer (in vivo) distal femora using a clinical CBCT scanner, and for reference, also using a conventional CT scanner. Our results indicated strong linear correlations between the vBMD values measured with the CT and CBCT scanners , however, absolute vBMD values were dependent on the scanner in use. Further, the differences between the vBMDs of cortical bone, trabecular bone and subchondral bone were similar and independent of the scanner. The present results indicate that vBMD values might not be directly comparable between different instruments. However, based on our present and previous results, we propose that, for OA diagnostics, clinical CBCT enables not only quantitative analysis of articular cartilage but also subchondral bone vBMD. Quantitative information on both cartilage and subchondral bone could be beneficial in OA diagnostics.
对比剂增强锥形束计算机断层扫描(CE-CBCT)是一种能够以小剂量辐射实现高分辨率体内成像的技术,已成功应用于软骨退变(即骨关节炎(OA))的临床诊断。作为一种 X 射线技术,CE-CBCT 还可能检测到软骨下骨(体积骨密度,vBMD)的矿物质密度变化,这是 OA 的特征。然而,由于 CBCT 图像的信噪比和对比度有限,其用于密度测量的可行性尚不清楚。在本研究中,我们创建了临床适用的羟基磷灰石体模,并使用临床 CBCT 扫描仪确定了 10 具尸体(离体)和 10 名志愿者(体内)的股骨远端皮质骨、小梁骨、软骨下小梁骨和软骨下板的 vBMD,作为参考,我们还使用了传统的 CT 扫描仪。我们的结果表明,使用 CT 和 CBCT 扫描仪测量的 vBMD 值之间存在很强的线性相关性,然而,绝对 vBMD 值取决于所使用的扫描仪。此外,皮质骨、小梁骨和软骨下骨的 vBMD 之间的差异相似,与扫描仪无关。目前的结果表明,不同仪器之间的 vBMD 值可能无法直接比较。然而,基于我们目前和以前的结果,我们提出,对于 OA 诊断,临床 CBCT 不仅能够定量分析关节软骨,还能够定量分析软骨下骨 vBMD。软骨和软骨下骨的定量信息可能有助于 OA 诊断。