用于舒张期反搏器的肌肉动力循环辅助装置。

Muscle powered circulatory assist device for diastolic counterpulsator.

作者信息

Novoa R, Jacobs G, Sakakibara N, Chen J F, Davies C, Cosgrove D M, Golding L R, Nosé Y, Loop F D

机构信息

Department of Thoracic and Cardiovascular Surgery, Cleveland Clinic Foundation, OH 44195.

出版信息

ASAIO Trans. 1989 Jul-Sep;35(3):408-11. doi: 10.1097/00002480-198907000-00076.

Abstract

A diastolic counterpulsator that uses either skeletal muscle or pneumatic actuation was developed. The unit is positioned between the latissimus dorsi and the chest wall, without interference with collateral blood supply, and is connected in series with the descending aorta. The system was able to generate stroke volumes between 52 and 16 ccs against pressures of 60 and 140 mmHg, respectively. Stroke work at 200 msec stimulation averaged 2.8 X 10(6) ergs. Power output at an afterload of 100 mmHg, and at a rate of 60 bpm, was 0.51 W. Back-up pneumatic actuation provided by an intraaortic balloon pump resulted in a 46% increase in the endocardial viability ratio (EVR).

摘要

研发出了一种使用骨骼肌或气动驱动的舒张期反搏器。该装置位于背阔肌和胸壁之间,不干扰侧支血液供应,并与降主动脉串联。该系统能够分别在60和140 mmHg的压力下产生52至16 cc的每搏输出量。在200毫秒刺激下的每搏功平均为2.8×10(6)尔格。在100 mmHg后负荷和60次/分钟的心率下,功率输出为0.51瓦。由主动脉内球囊泵提供的备用气动驱动使心内膜活力比率(EVR)提高了46%。