Zantek N D, Hsu P, Meijer P, Smock K J, Plumhoff E A, Refaai M A, Van Cott E M
Department of Laboratory Medicine and Pathology, University of Minnesota, Minneapolis, MN, USA.
Department of Pathology/Laboratory Medicine, Hofstra North Shore-Long Island Jewish School of Medicine, Lake Success, NY, USA.
Int J Lab Hematol. 2015 May;37 Suppl 1:99-107. doi: 10.1111/ijlh.12359.
The performance of factor XI activity (FXI) by laboratories in the North American Specialized Coagulation Laboratory Association proficiency testing program was analyzed.
Over 10 years (2003-2013), 80 samples were distributed; 33-55 laboratories participated per exercise providing 3833 total responses. Analysis was performed on numeric results and qualitative classification of results.
The sample FXI levels ranged from 3.8 to 154.0 IU/dL. The overall interlaboratory average coefficient of variation (CV%) was 17.5%; the CV was higher for a sample with low (3.8 IU/dL) FXI. Results were correctly classified as abnormal (100%) for a sample with 3.8 IU/dL FXI and normal/borderline normal (97.7%) for 45 samples with 80 to < 140 IU/dL FXI. The classification was heterogeneous for samples with FXI of 50 to < 80 IU/dL. Six specimens were repeat-tested from 2007 to 2013. The mean FXI was not significantly different in laboratories using the same method on both exercises, suggesting good intralaboratory precision over time. Univariate analysis of data from 2011 to 2012 did not find a consistent significant difference among the activators, analyzers, calibrators, and FXI-deficient plasmas.
Laboratories generally performed well in assessment of FXI based on interlaboratory precision when FXI >30 IU/dL and on classification of samples with very low or normal FXI.
分析了北美专业凝血实验室协会能力验证计划中各实验室对凝血因子 XI 活性(FXI)的检测情况。
在超过 10 年(2003 - 2013 年)的时间里,共分发了 80 个样本;每次检测有 33 - 55 个实验室参与,总共提供了 3833 份检测结果。对数值结果和结果的定性分类进行了分析。
样本的 FXI 水平在 3.8 至 154.0 IU/dL 之间。实验室间总体平均变异系数(CV%)为 17.5%;FXI 水平低(3.8 IU/dL)的样本变异系数更高。FXI 为 3.8 IU/dL 的样本结果被正确分类为异常(100%),FXI 为 80 至<140 IU/dL 的 45 个样本结果被正确分类为正常/临界正常(97.7%)。FXI 为 50 至<80 IU/dL 的样本分类存在异质性。2007 年至 2013 年期间对 6 个样本进行了重复检测。在两次检测中使用相同方法的实验室,其平均 FXI 无显著差异,表明实验室内部长期精度良好。对 2011 年至 2012 年的数据进行单因素分析,未发现激活剂、分析仪、校准品和 FXI 缺乏血浆之间存在一致的显著差异。
当 FXI>30 IU/dL 时,实验室在基于实验室间精度评估 FXI 以及对 FXI 极低或正常的样本进行分类方面总体表现良好。