Suppr超能文献

与传统导尿法相比,三维膀胱超声检查测定残余尿量的准确性。

The accuracy of three-dimensional bladder ultrasonography in determining the residual urinary volume compared with conventional catheterisation.

作者信息

Jalbani Imran K, Ather M Hammad

机构信息

Department of Surgery, Aga Khan University, Karachi, Pakistan.

出版信息

Arab J Urol. 2014 Sep;12(3):209-13. doi: 10.1016/j.aju.2014.05.001. Epub 2014 Jul 29.

Abstract

OBJECTIVE

To determine the accuracy of three-dimensional bladder ultrasonography (US, using the BVI 3000, Verathon, WA, USA) for determining the residual urinary volume, compared with the conventional catheterisation method.

PATIENTS AND METHODS

We conducted a cross-sectional study at day-care unit of a University hospital after obtaining approval from the Ethics Review Committee of the hospital. Thirty-four patients with lower urinary tract symptoms requiring cystometrography were included. The postvoid volume was measured by bladder US, with three readings taken, and then patient was catheterised using a 12-F Nelaton catheter to measure the urinary volume. The mean of the three readings was compared with the catheterisation volume.

RESULTS

The mean (SD) urinary volumes by US and catheterisation were 261 (186) and 260 (175) mL, respectively, and the correlation (r (2)) was 0.97. There was no effect of age, gender or body mass index on the accuracy of bladder US, which was accurate even when the urinary volume was ⩽100 mL.

CONCLUSION

The bladder US estimate is as accurate as catheterisation for determining the postvoid residual urinary volume. Its accuracy was also comparable when the urinary volume is <100 mL, and there was no significant effect of age, gender and body mass index. This system could replace the more invasive catheterisation, and with excellent accuracy.

摘要

目的

与传统导尿法相比,确定三维膀胱超声检查(使用美国华盛顿州Verathon公司的BVI 3000型超声仪)测定残余尿量的准确性。

患者与方法

在获得医院伦理审查委员会批准后,我们在一家大学医院的日间病房进行了一项横断面研究。纳入34例需要进行膀胱压力测定的下尿路症状患者。通过膀胱超声测量排尿后膀胱容量,测量3次读数,然后使用12F Nelaton导管对患者进行导尿以测量尿量。将3次读数的平均值与导尿量进行比较。

结果

超声检查和导尿法测得的平均(标准差)尿量分别为261(186)mL和260(175)mL,相关性(r²)为0.97。年龄、性别或体重指数对膀胱超声检查的准确性没有影响,即使尿量≤100 mL时超声检查也很准确。

结论

膀胱超声检查在测定排尿后残余尿量方面与导尿法一样准确。当尿量<100 mL时,其准确性也相当,且年龄、性别和体重指数对其无显著影响。该系统可以替代侵入性更强的导尿法,且准确性极佳。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/8eb3/4435769/641529bb5f9e/gr1.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验