Kochhar R, Saxena R, Singh R, Mehta S K
Department of Gastroenterology, Postgraduate Institute of Medical Education and Research, Chandigarh, India.
J Emerg Med. 1989 Sep-Oct;7(5):469-70. doi: 10.1016/0736-4679(89)90147-9.
Impaction of the Sengstaken-Blakemore tube may occur due to incorrect placement of clamps, biting of the outlet channel(s) or inadvertent injection of particulate matter at the time of inflation of the balloon(s). Endoscopic deflation of the impacted balloon using a pointed object is an easy remedy. This report highlights the pathogenesis of this avoidable complication in one patient.
三腔二囊管堵塞可能由于夹子放置不当、咬损出口通道或在气囊充气时意外注入颗粒物所致。使用尖锐物体在内镜下使堵塞的气囊放气是一种简便的补救方法。本报告着重介绍了一名患者发生这种可避免并发症的发病机制。