Clemente-Gutiérrez Francisco, Pérez-Vara Consuelo
Hospital Central de la Defensa "Gomez Ulla".
J Appl Clin Med Phys. 2015 Mar 8;16(2):5190. doi: 10.1120/jacmp.v16i2.5190.
A pretreatment quality assurance program for volumetric techniques should include redundant calculations and measurement-based verifications. The patient-specific quality assurance process must be based in clinically relevant metrics. The aim of this study was to show the commission, clinical implementation, and comparison of two systems that allow performing a 3D redundant dose calculation. In addition, one of them is capable of reconstructing the dose on patient anatomy from measurements taken with a 2D ion chamber array. Both systems were compared in terms of reference calibration data (absolute dose, output factors, percentage depth-dose curves, and profiles). Results were in good agreement for absolute dose values (discrepancies were below 0.5%) and output factors (mean differences were below 1%). Maximum mean discrepancies were located between 10 and 20 cm of depth for PDDs (-2.7%) and in the penumbra region for profiles (mean DTA of 1.5 mm). Validation of the systems was performed by comparing point-dose measurements with values obtained by the two systems for static, dynamic fields from AAPM TG-119 report, and 12 real VMAT plans for different anatomical sites (differences better than 1.2%). Comparisons between measurements taken with a 2D ion chamber array and results obtained by both systems for real VMAT plans were also performed (mean global gamma passing rates better than 87.0% and 97.9% for the 2%/2 mm and 3%/3 mm criteria). Clinical implementation of the systems was evaluated by comparing dose-volume parameters for all TG-119 tests and real VMAT plans with TPS values (mean differences were below 1%). In addition, comparisons between dose distributions calculated by TPS and those extracted by the two systems for real VMAT plans were also performed (mean global gamma passing rates better than 86.0% and 93.0% for the 2%/2 mm and 3%/ 3 mm criteria). The clinical use of both systems was successfully evaluated.
容积技术的预处理质量保证程序应包括冗余计算和基于测量的验证。针对特定患者的质量保证过程必须基于临床相关指标。本研究的目的是展示两种能够进行三维冗余剂量计算的系统的调试、临床应用及比较。此外,其中一个系统能够根据二维电离室阵列的测量结果在患者解剖结构上重建剂量。对这两个系统在参考校准数据(绝对剂量、输出因子、百分深度剂量曲线和剂量分布)方面进行了比较。绝对剂量值(差异低于0.5%)和输出因子(平均差异低于1%)的结果吻合良好。百分深度剂量曲线在深度10至20厘米之间的最大平均差异为-2.7%,剂量分布在半影区的最大平均差异为1.5毫米。通过将点剂量测量结果与这两个系统针对AAPM TG-119报告中的静态、动态射野以及12个不同解剖部位的真实容积调强放疗计划所获得的值进行比较,对系统进行了验证(差异优于1.2%)。还对二维电离室阵列的测量结果与这两个系统针对真实容积调强放疗计划所获得的结果进行了比较(对于2%/2毫米和3%/3毫米标准,平均全局γ通过率分别优于87.0%和97.9%)。通过将所有TG-119测试和真实容积调强放疗计划的剂量体积参数与治疗计划系统的值进行比较,对系统的临床应用进行了评估(平均差异低于1%)。此外,还对治疗计划系统计算的剂量分布与这两个系统针对真实容积调强放疗计划提取的剂量分布进行了比较(对于2%/2毫米和3%/3毫米标准,平均全局γ通过率分别优于86.0%和93.0%)。成功评估了这两个系统的临床应用情况。