Transtracheal ventilation with a novel ejector-based device (Ventrain) in open, partly obstructed, or totally closed upper airways in pigs.

作者信息

Paxian M, Preussler N P, Reinz T, Schlueter A, Gottschall R

机构信息

Department of Anaesthesiology and Intensive Care Medicine, Jena University Hospital, Erlanger Allee 101, 07747 Jena, Germany

Department of Anaesthesiology and Intensive Care Medicine, Jena University Hospital, Erlanger Allee 101, 07747 Jena, Germany.

出版信息

Br J Anaesth. 2015 Aug;115(2):308-16. doi: 10.1093/bja/aev200. Epub 2015 Jun 25.

Abstract

BACKGROUND

Transtracheal access and subsequent jet ventilation are among the last options in a 'cannot intubate-cannot oxygenate' scenario. These interventions may lead to hypercapnia, barotrauma, and haemodynamic failure in the event of an obstructed upper airway. The aim of the present study was to evaluate the efficacy and the haemodynamic effects of the Ventrain, a manually operated ventilation device that provides expiratory ventilation assistance. Transtracheal ventilation was carried out with the Ventrain in different airway scenarios in live pigs, and its performance was compared with a conventional jet ventilator.

METHODS

Pigs with open, partly obstructed, or completely closed upper airways were transtracheally ventilated either with the Ventrain or by conventional jet ventilation. Airway pressures, haemodynamic parameters, and blood gases obtained in the different settings were compared.

RESULTS

Mean (SD) alveolar minute ventilation as reflected by arterial partial pressure of CO2 was superior with the Ventrain in partly obstructed airways after 6 min in comparison with traditional manual jet ventilation [4.7 (0.19) compared with 7.1 (0.37) kPa], and this was also the case in all simulated airway conditions. At the same time, peak airway pressures were significantly lower and haemodynamic parameters were altered to a lesser extent with the Ventrain.

CONCLUSIONS

The results of this study suggest that the Ventrain device can ensure sufficient oxygenation and ventilation through a small-bore transtracheal catheter when the airway is open, partly obstructed, or completely closed. Minute ventilation and avoidance of high airway pressures were superior in comparison with traditional hand-triggered jet ventilation, particularly in the event of complete upper airway obstruction.

摘要

背景

在“无法插管-无法给氧”的情况下,经气管穿刺通气及随后的喷射通气是最后的选择之一。如果上气道梗阻,这些干预措施可能会导致高碳酸血症、气压伤和血流动力学衰竭。本研究的目的是评估Ventrain(一种提供呼气通气辅助的手动通气设备)的疗效和血流动力学效应。在活猪的不同气道情况下使用Ventrain进行经气管通气,并将其性能与传统喷射呼吸机进行比较。

方法

对上气道开放、部分梗阻或完全闭合的猪,分别使用Ventrain或传统喷射通气进行经气管通气。比较不同设置下获得的气道压力、血流动力学参数和血气指标。

结果

与传统手动喷射通气相比,在部分梗阻气道中,Ventrain在6分钟后由动脉二氧化碳分压反映的平均(标准差)肺泡分钟通气量更高[4.7(0.19)kPa 对比 7.1(0.37)kPa],在所有模拟气道条件下均如此。同时,Ventrain的气道峰值压力显著更低,血流动力学参数改变程度更小。

结论

本研究结果表明,当气道开放、部分梗阻或完全闭合时,Ventrain设备可通过细径经气管导管确保充足的氧合和通气。与传统手动触发喷射通气相比,分钟通气量及避免高气道压力方面更优,尤其是在上气道完全梗阻的情况下。