Calibration of Clinical Audio Recording and Analysis Systems for Sound Intensity Measurement.

作者信息

Maryn Youri, Zarowski Andrzej

出版信息

Am J Speech Lang Pathol. 2015 Nov;24(4):608-18. doi: 10.1044/2015_AJSLP-14-0082.

Abstract

PURPOSE

Sound intensity is an important acoustic feature of voice/speech signals. Yet recordings are performed with different microphone, amplifier, and computer configurations, and it is therefore crucial to calibrate sound intensity measures of clinical audio recording and analysis systems on the basis of output of a sound-level meter. This study was designed to evaluate feasibility, validity, and accuracy of calibration methods, including audiometric speech noise signals and human voice signals under typical speech conditions.

METHOD

Calibration consisted of 3 comparisons between data from 29 measurement microphone-and-computer systems and data from the sound-level meter: signal-specific comparison with audiometric speech noise at 5 levels, signal-specific comparison with natural voice at 3 levels, and cross-signal comparison with natural voice at 3 levels. Intensity measures from recording systems were then linearly converted into calibrated data on the basis of these comparisons, and validity and accuracy of calibrated sound intensity were investigated.

RESULTS

Very strong correlations and quasisimilarity were found between calibrated data and sound-level meter data across calibration methods and recording systems.

CONCLUSIONS

Calibration of clinical sound intensity measures according to this method is feasible, valid, accurate, and representative for a heterogeneous set of microphones and data acquisition systems in real-life circumstances with distinct noise contexts.