Suppr超能文献

一种从人血清中支持液液萃取视黄醇和α-生育酚并通过液相色谱串联质谱法同时定量的新型高通量方法。

A novel high-throughput method for supported liquid extraction of retinol and alpha-tocopherol from human serum and simultaneous quantitation by liquid chromatography tandem mass spectrometry.

作者信息

Hinchliffe Edward, Rudge James, Reed Paul

机构信息

Department of Clinical Biochemistry, Salford Royal Hospital, Salford, UK

Phenomenex UK Ltd, Macclesfield, UK.

出版信息

Ann Clin Biochem. 2016 Jul;53(Pt 4):434-45. doi: 10.1177/0004563215596024. Epub 2015 Jul 20.

Abstract

BACKGROUND

Measurement of vitamin A (retinol) and E (alpha-tocopherol) in UK clinical laboratories is currently performed exclusively by high-performance liquid chromatography with ultraviolet detection. We investigated whether retinol and alpha-tocopherol could be measured simultaneously by liquid chromatography tandem mass spectrometry.

METHODS

Serum samples (100 μL) were extracted using Isolute + Supported Liquid Extraction plates. Chromatography was performed on a Phenomenex Kinetex Biphenyl 2.6 μm, 50 × 2.1 mm column, and liquid chromatography tandem mass spectrometry on a Waters Acquity TQD. Injection-to-injection time was 4.3 min. The assay was validated according to published guidelines. Patient samples were used to compare liquid chromatography tandem mass spectrometry and high-performance liquid chromatography with ultraviolet detection methods.

RESULTS

For retinol and alpha-tocopherol, respectively, the assay was linear up to 6.0 and 80.0 μmol/L, and lower limit of quantification was 0.07 and 0.26 μmol/L. Intra and interassay imprecision were within desirable analytical specifications. Analysis of quality control material aligned to NIST SRM 968e, and relative spiked recovery from human serum, both yielded results within 15% of target values. Method comparison with high-performance liquid chromatography with ultraviolet detection methodology demonstrated a negative bias for retinol and alpha-tocopherol by the liquid chromatography tandem mass spectrometry method. Analysis of United Kingdom National External Quality Assurance Scheme samples yielded mean bias from the target value of +3.0% for retinol and -11.2% for alpha-tocopherol.

CONCLUSIONS

We have developed a novel, high-throughput method for extraction of retinol and alpha-tocopherol from human serum followed by simultaneous quantitation by liquid chromatography tandem mass spectrometry. The method offers a rapid, sensitive, specific and cost-effective alternative to high-performance liquid chromatography with ultraviolet detection methodology, and is suitable for routine clinical monitoring of patients predisposed to fat-soluble vitamin malabsorption.

摘要

背景

目前英国临床实验室中维生素A(视黄醇)和维生素E(α-生育酚)的检测完全通过带紫外检测的高效液相色谱法进行。我们研究了视黄醇和α-生育酚是否可以通过液相色谱串联质谱法同时进行检测。

方法

使用Isolute + 支持液液萃取板提取血清样本(100 μL)。在Phenomenex Kinetex联苯2.6 μm、50 × 2.1 mm柱上进行色谱分析,并在Waters Acquity TQD上进行液相色谱串联质谱分析。进样间隔时间为4.3分钟。该检测方法按照已发表的指南进行验证。使用患者样本比较液相色谱串联质谱法和带紫外检测的高效液相色谱法。

结果

视黄醇和α-生育酚的检测方法线性范围分别高达6.0和80.0 μmol/L,定量下限分别为0.07和0.26 μmol/L。批内和批间不精密度均在理想的分析规格范围内。对与NIST SRM 968e校准的质控物质进行分析,以及从人血清中加标后的相对回收率,结果均在目标值的15%以内。与带紫外检测的高效液相色谱法进行方法比较表明,液相色谱串联质谱法对视黄醇和α-生育酚存在负偏差。对英国国家外部质量保证计划样本的分析得出,视黄醇相对于目标值的平均偏差为 +3.0%,α-生育酚为 -11.2%。

结论

我们开发了一种新颖的高通量方法,用于从人血清中提取视黄醇和α-生育酚,随后通过液相色谱串联质谱法进行同时定量。该方法为带紫外检测的高效液相色谱法提供了一种快速、灵敏、特异且经济高效的替代方法,适用于对易发生脂溶性维生素吸收不良患者的常规临床监测。