Repeatability of Retinal Macular Thickness Measurements in Patients with Clinically Significant Macular Edema Using Two Different Scanning Protocols of Spectralis Optical Coherence Tomography.

作者信息

Fiore Tito, Lupidi Marco, Androudi Sofia, Giansanti Fabrizio, Fruttini Daniela, Cagini Carlo

机构信息

Department of Ophthalmology, Eye Clinic, University of Perugia, Perugia, Italy.

出版信息

Ophthalmologica. 2015;234(3):167-71. doi: 10.1159/000435841. Epub 2015 Jul 21.

Abstract

OBJECTIVE

To determine the repeatability of Spectralis optical coherence tomography (OCT) retinal thickness measurements in diabetic patients with clinically significant macular edema (CSME) using two different scanning protocols.

METHODS

Seventy-one eyes of 71 diabetic patients with CSME were included in the study. Coefficients of repeatability and intrasession variation coefficients were tested with 20 × 15 degree raster scans consisting of 19 or 37 high-resolution line scans (15 or 8 frames per scan, respectively) that were repeated 2 times by 1 experienced examiner. The first scan was set as the reference scan; the second scan was the follow-up scan and was performed with the use of the follow-up mode.

RESULTS

The mean and standard deviation for the central foveal subfield (CSF) using the first scanning method was 404 ± 88 μm, while it was 399 ± 86 μm using the second protocol, which was not statistically significantly different (p = 0.35, paired test). Particularly examining the CSF, the coefficient of repeatability was 1.48 (6.00 µm) and 1.49 (5.95 µm) for the 19- and the 37-B-scan acquisition, respectively, showing a nonstatistically significant difference (p < 0.001).

CONCLUSIONS

Retinal thickness measurements in diabetic patients with CSME are repeatable using both scanning protocols (19 or 37 B-scans) with Spectralis OCT. The repeatability of the retinal thickness measurement does not improve by increasing the number of B-scans from 19 to 37.

摘要

目的

使用两种不同的扫描方案,确定光谱域光学相干断层扫描(OCT)对患有临床显著性黄斑水肿(CSME)的糖尿病患者视网膜厚度测量的可重复性。

方法

71例患有CSME的糖尿病患者的71只眼纳入本研究。由1名经验丰富的检查者对20×15度光栅扫描(分别由19或37条高分辨率线扫描组成,每次扫描分别为15或8帧)进行2次重复扫描,测试可重复性系数和扫描期间内变异系数。第一次扫描设为参考扫描;第二次扫描为随访扫描,采用随访模式进行。

结果

使用第一种扫描方法时,中心凹子区域(CSF)的平均值和标准差为404±88μm,而使用第二种方案时为399±86μm,差异无统计学意义(配对检验,p = 0.35)。特别检查CSF时,19次和37次B扫描采集的可重复性系数分别为1.48(6.00μm)和1.49(5.95μm),差异无统计学意义(p < 0.001)。

结论

使用光谱域OCT的两种扫描方案(19次或37次B扫描)对患有CSME的糖尿病患者进行视网膜厚度测量均具有可重复性。将B扫描次数从19次增加到37次并不能提高视网膜厚度测量的可重复性。