Sasaki K, Isimura Y, Fujii K, Wake K, Watanabe S, Kojima M, Suga R, Hashimoto O
National Institute of Information and Communications Technology, Koganei, Tokyo 184-8795, Japan.
Phys Med Biol. 2015 Aug 21;60(16):6273-88. doi: 10.1088/0031-9155/60/16/6273. Epub 2015 Aug 3.
Measurement of the dielectric properties of ocular tissues up to 110 GHz was performed by the coaxial probe method. A coaxial sensor was fabricated to allow the measurement of small amounts of biological tissues. Four-standard calibration was applied in the dielectric property measurement to obtain more accurate data than that obtained with conventional three-standard calibration, especially at high frequencies. Novel data of the dielectric properties of several ocular tissues are presented and compared with data from the de facto database.
采用同轴探头法对高达110 GHz的眼组织介电特性进行了测量。制作了一种同轴传感器,用于测量少量生物组织。在介电特性测量中采用四标准校准,以获得比传统三标准校准更准确的数据,尤其是在高频时。给出了几种眼组织介电特性的新数据,并与实际数据库中的数据进行了比较。