自动靶控麻醉机吸入药物使用显示的准确性

Accuracy of inhaled agent usage displays of automated target control anesthesia machines.

作者信息

Dehouwer Alexander, Carette Rik, De Ridder Simon, De Wolf Andre M, Hendrickx Jan F A

机构信息

Department of Anesthesiology, Intensive Care and Pain Therapy, OLV Hospital, Moorselbaan 164, 9300, Aalst, Belgium.

Department of Anesthesiology, Feinberg School of Medicine, Northwestern University, Chicago, IL, USA.

出版信息

J Clin Monit Comput. 2016 Oct;30(5):539-43. doi: 10.1007/s10877-015-9746-z. Epub 2015 Aug 7.

Abstract

Automated low flow anesthesia machines report how much inhaled anesthetic agent has been used for each anesthetic. We compared these reported values with the amount of agent that had disappeared by weighing the vaporizer/injectors before and after each anesthetic. The vaporizers/injectors of the Aisys, Zeus and FLOW-i were weighed with a high precision weighing scale before and after anesthesia with either desflurane in O2/air or sevoflurane in O2/N2O. These values were compared with the values reported by the cumulative agent use display tools of the respective anesthesia machines using a linear curve fit. Twenty-five measurements were performed in each group, except for the sevoflurane data with the Aisys that were available from another study (87 pairs). We also determined the amount lost by inserting and removing the vaporizers/injectors or by performing a machine checkout, corrected the measured amounts for these artifacts and repeated the linear fits. The average amount of sevoflurane and desflurane wasted by inserting and removing the cassette for the Aisys, Zeus, and FLOW-i were 0.21, 0.12, and 0.04 mL and 0.12, 0.61, and 1.13 mL liquid agent, respectively. The average amount of sevoflurane and desflurane wasted by the machine checkout with the Aisys, Zeus, and FLOW-i were 1.78, 0.21, and 1.67 mL and 2.39, 0.67, and 4.19 mL, respectively. Performance error of the displayed amount of agent use remained within 10 % of the weighed amount, expect for amounts less than 3 mL sevofurane with the FLOW-i and less than 20 mL desflurane with the Aisys and FLOW-i. Cumulative agent usage displayed by the Aisys, Zeus, and FLOW-i is within 10 % of the measured consumption, except for low consumption cases (<3 mL sevoflurane, <20 mL desflurane). The differences may be due to either measurement error or cumulative agent display error. The current results can help the researchers decide whether the displayed amounts are accurate enough for their study purposes. The extent to which these discrepancies differ between different units of the same machine remains unstudied.

摘要

自动低流量麻醉机可报告每次麻醉所使用的吸入麻醉剂的量。我们通过在每次麻醉前后对蒸发器/注射器进行称重,来比较这些报告值与消失的麻醉剂的量。在使用氧气/空气混合气体中的地氟烷或氧气/氧化亚氮混合气体中的七氟烷进行麻醉前后,使用高精度秤对Aisys、Zeus和FLOW-i麻醉机的蒸发器/注射器进行称重。使用线性曲线拟合,将这些值与相应麻醉机的累积麻醉剂使用显示工具报告的值进行比较。每组进行25次测量,但Aisys麻醉机的七氟烷数据来自另一项研究(87对)。我们还通过插入和移除蒸发器/注射器或进行机器检查来确定损失的量,对这些假象的测量量进行校正,然后重复线性拟合。对于Aisys、Zeus和FLOW-i麻醉机,插入和移除药盒浪费的七氟烷和地氟烷的平均量分别为0.21、0.12和0.04 mL以及0.12、0.61和1.13 mL液体麻醉剂。对于Aisys、Zeus和FLOW-i麻醉机,机器检查浪费的七氟烷和地氟烷的平均量分别为1.78、0.21和1.67 mL以及2.39、0.67和4.19 mL。除了FLOW-i麻醉机使用量小于3 mL的七氟烷以及Aisys和FLOW-i麻醉机使用量小于20 mL的地氟烷外,显示的麻醉剂使用量的性能误差仍在称重值的10%以内。Aisys、Zeus和FLOW-i麻醉机显示的累积麻醉剂使用量在测量消耗量的10%以内,但低消耗量情况(<3 mL七氟烷,<20 mL地氟烷)除外。这些差异可能是由于测量误差或累积麻醉剂显示误差。当前结果可帮助研究人员确定显示的量对于其研究目的是否足够准确。同一台机器不同单元之间这些差异的程度仍未得到研究。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验