Effects of Pulsatile Fatigue on In Situ Antegrade Fenestrated Polyester Stent Grafts Deployed in a Patient-Specific Phantom Model of Juxtarenal Aortic Aneurysm.

作者信息

Ruthrauff Alicia A, King Martin W, Soulez Gilles, Tan Kong T, Crawford Sean A, Roche-Nagle Graham, Cloutier Guy, Tse Leonard W

机构信息

College of Textiles, North Carolina State University, Raleigh, North Carolina.

College of Textiles, North Carolina State University, Raleigh, North Carolina; College of Textiles, Donghua University, Shanghai, China.

出版信息

J Vasc Interv Radiol. 2015 Oct;26(10):1551-8. doi: 10.1016/j.jvir.2015.06.038. Epub 2015 Aug 4.

Abstract

PURPOSE

To evaluate the effects of in situ fenestration on the fabric of stent grafts deployed in a patient-specific phantom of a juxtarenal abdominal aortic aneurysm.

MATERIALS AND METHODS

Four patient-specific juxtarenal abdominal aortic aneurysm polyurethane models were created, and bifurcated Zenith (Cook, Inc, Bloomington, Indiana) and Endurant (Medtronic, Minneapolis, Minneapolis) endografts were deployed into the models, covering the renal arteries. Antegrade in situ fenestration was carried out with radiofrequency puncture followed by balloon dilation with either conventional or cutting balloons. Renal covered stents were deployed and flared. Specimens were mounted onto an accelerated fatigue tester for 40M cycles (1 patient life-year), and evaluated with microscopy, caliper measurements, and fabric counts.

RESULTS

Cutting balloons resulted in more fabric fraying. None of the fenestrations grew beyond the targeted 6-mm diameter despite accelerated fatigue. Fluoroscopic images demonstrated a very prominent waist of the renal fenestration in the Cook device when a conventional balloon was used compared with a cutting balloon. The average fenestration diameter for the Cook device was only 3.1 mm with the conventional balloon compared with 4.8 mm with the cutting balloon. The average fenestration diameter for the Medtronic device was 3.8 mm with the conventional balloon compared with 5.1 mm with the cutting balloon. The fabric counts suggested crowding of yarns around the fenestrations with conventional balloons but less with cutting balloons.

CONCLUSIONS

This experimental work suggests that the size of in situ renal fenestrations does not expand beyond the target diameter despite cyclic fatigue. Although the small number of devices tested and selected aortorenal anatomy in this study may limit conclusions, textile analysis suggests that cutting balloons should be used for the Cook Zenith device, whereas conventional balloons should be used for the Medtronic Endurant device when performing in situ fenestration.