To validate a self-expanding transcatheter valve for off-pump transatrial mitral valve-in-ring (VIR) implantation via a left thoracotomy.
Mitral valve annuloplasty was performed via sternotomy during cardiopulmonary bypass on 9 pigs. After successful weaning from extracorporal circulation, the custom-made, self-expanding transcatheter VIR device was deployed under fluoroscopic guidance within the annuloplasty ring via a left thoracotomy. Hemodynamic data before and after the implantation were recorded. Mitral annulus diameter and valve area were measured by echocardiography. Transvalvular and left-ventricular outflow-tract pressure gradient were measured invasively.
Eight successful implantations were performed. Implantation failed in 1 pig because of difficulty with technical delivery of the sheath. Mean transatrial procedure time was 12.6 ± 1.7 min. Hemodynamic status during transatrial implantation was stable, and differences were not statistically significant. Mean mitral annulus diameter and mean mitral orifice area were 2.32 ± 0.2 and 3.84 ± 0.55 cm2, respectively. Mild regurgitation was detected in 7 animals and moderate regurgitation in 1. Mean gradients were 6.1 ± 5.0 mm Hg across the device. Postmortem examination confirmed adequate positioning of devices within the annuloplasty ring.
This custom-made transcatheter device allows for safe and reproducible off-pump transatrial mitral VIR implantations. Transatrial access is a promising route to facilitate VIR implantations. Our custom-made stent-valve may be suitable for VIR procedures.
验证一种用于经左胸小切口非体外循环下经房间隔二尖瓣环内(VIR)植入的自膨胀经导管瓣膜。
对9头猪在体外循环下经胸骨正中切口行二尖瓣环成形术。成功脱离体外循环后,在透视引导下经左胸小切口将定制的自膨胀经导管VIR装置植入二尖瓣环成形环内。记录植入前后的血流动力学数据。通过超声心动图测量二尖瓣环直径和瓣膜面积。有创测量跨瓣和左心室流出道压力梯度。
成功植入8例。1头猪因鞘管技术输送困难植入失败。平均经房间隔手术时间为12.6±1.7分钟。经房间隔植入期间血流动力学状态稳定,差异无统计学意义。平均二尖瓣环直径和平均二尖瓣口面积分别为2.32±0.2和3.84±0.55平方厘米。7只动物检测到轻度反流,1只动物检测到中度反流。装置跨瓣平均压差为6.1±5.0毫米汞柱。尸检证实装置在二尖瓣环成形环内定位良好。
这种定制的经导管装置允许安全且可重复地进行非体外循环下经房间隔二尖瓣VIR植入。经房间隔途径是促进VIR植入的一种有前景的途径。我们定制的支架瓣膜可能适用于VIR手术。