An animal model has been designed to assess the feasibility of off-pump mitral valve replacement using valved stents.
Glutaraldehyde-preserved homograft was sutured inside a prosthetic tube (Dacron). Then, two self-expandable nitinol Z-stents were sutured on the external surface of the prosthesis in such a way to create two self-expanding crowns for fixation. In adult pigs and under general anesthesia, the left atrium was exposed through a left thoracotomy and atrio-ventricular roadmapping was performed with intravascular ultrasound (IVUS) and fluoroscopy. The double-crowned valved stents were loaded into a delivery sheath. The sheath was then introduced into the left atrium and the valved stents was deployed in mitral position in such a way that the part in between the two stents was at the level of the mitral annulus. Intracardiac Unltrasound (ICUS) was used to assess the valve function. Hemodynamic parameters were gathered as well. Animal survived for no more than 3h after the valve deployment and gross anatomy examination of the left heart was carried out.
The mean height of the valved stents was 29.4+/-0.2 mm, with an internal diameter of 20.4+/-1.0mm, and an external diameter of 25.5+/-0.8 mm. The procedure was successfully carried out in eight animals. In vivo evaluation showed a native mitral annulus diameter of 24.9+/-0.6 mm, and a mean mitral valve area of 421.4+/-17.5 mm2. ICUS showed a mild mitral regurgitation in three out of eight animals. Mean pressure gradient across the valved stents was 2.6+/-3.1 mmHg. Mean pressure gradient across the left ventricular outflow tract (LVOT) was 6.6+/-5.2 mmHg. The mean survival time was 97.5+/-56.3 min (survival time range was 40-180 min). One animal died due to the occlusion of the LVOT because of valved stents displacement. Postmortem evaluation confirmed correct positioning of the valved stent in the mitral position in seven out of eight animals. No atrial or ventricular lesions due to the valved stents were found.
Off-pump implantation of a self-expandable valved stent in the mitral position is technically feasible. Further studies will assess if this procedure is also feasible in humans.
设计一种动物模型,以评估使用带瓣支架进行非体外循环二尖瓣置换术的可行性。
将戊二醛保存的同种异体移植物缝合在人工血管(涤纶)内。然后,将两个自膨胀镍钛合金Z形支架以形成两个用于固定的自膨胀冠的方式缝合在人工血管的外表面。在成年猪全身麻醉下,通过左胸切口暴露左心房,并使用血管内超声(IVUS)和荧光透视进行房室路线图绘制。将双冠带瓣支架装入输送鞘管。然后将鞘管插入左心房,并将带瓣支架置于二尖瓣位置,使两个支架之间的部分位于二尖瓣环水平。使用心内超声(ICUS)评估瓣膜功能。同时收集血流动力学参数。瓣膜置入后动物存活不超过3小时,然后对左心进行大体解剖检查。
带瓣支架的平均高度为29.4±0.2mm,内径为20.4±1.0mm,外径为25.5±0.8mm。该手术在8只动物中成功实施。体内评估显示天然二尖瓣环直径为24.9±0.6mm,平均二尖瓣面积为421.4±17.5mm²。ICUS显示8只动物中有3只存在轻度二尖瓣反流。跨带瓣支架的平均压力梯度为2.6±3.1mmHg。跨左心室流出道(LVOT)的平均压力梯度为6.6±5.2mmHg。平均存活时间为97.5±56.3分钟(存活时间范围为40 - 180分钟)。1只动物因带瓣支架移位导致LVOT阻塞死亡。尸检评估证实8只动物中有7只带瓣支架在二尖瓣位置定位正确。未发现因带瓣支架导致的心房或心室病变。
在二尖瓣位置非体外循环植入自膨胀带瓣支架在技术上是可行的。进一步的研究将评估该手术在人类中是否也可行。