Hranitzky C
Dosimetry Laboratory Seibersdorf, Seibersdorf Labor GmbH, Seibersdorf 2444, Austria
Radiat Prot Dosimetry. 2016 Sep;170(1-4):27-30. doi: 10.1093/rpd/ncv384. Epub 2015 Aug 25.
The Dosimetry Laboratory Seibersdorf was the irradiation laboratory for the IC2014 EURADOS intercomparison for individual monitoring. One hundred and twelve whole-body dosimetry systems with 30 dosemeters per dosimetry system participated in the intercomparison. In total, 2240 personal dosemeters were irradiated in photon reference radiation fields on the ISO water slab phantom in terms of personal dose equivalent Hp(10) and Hp(0.07). Two nuclide irradiation systems ((137)Cs and (60)Co gamma radiation) and a 320-kV X-ray radiation unit were used. The irradiation plan consisted of nine irradiation set-ups with five different ISO/IEC photon radiation qualities (S-Cs, S-Co, RQR 7, W-80 and W-150) and two different angles of radiation incidence (0° and 60°). Reference dose equivalent values were in the range of ∼1-500 mSv. Reference values were based on air kerma rate traceable to primary standard calibrations and appropriate conversion coefficients resulting in total expanded measurement uncertainties <6 %.
赛伯斯多夫剂量测定实验室是用于IC2014 EURADOS个人监测比对的辐照实验室。112个全身剂量测定系统参与了此次比对,每个剂量测定系统配备30个剂量计。总共2240个个人剂量计在ISO水模体的光子参考辐射场中按照个人剂量当量Hp(10)和Hp(0.07)进行了辐照。使用了两个核素辐照系统((137)Cs和(60)Coγ辐射)以及一个320 kV X射线辐射装置。辐照计划包括九个辐照设置,具有五种不同的ISO/IEC光子辐射品质(S-Cs、S-Co、RQR 7、W-80和W-150)以及两种不同的辐射入射角(0°和60°)。参考剂量当量值范围约为1 - 500 mSv。参考值基于可溯源至基准标准校准的空气比释动能率以及适当的转换系数,从而使总扩展测量不确定度<6%。